background image

AC/DC Current Probe Models MR411 & MR521 

- 3 - 

Definition of Measurement Categories 

CAT II: 

For measurements performed on circuits directly connected to 
the electrical distribution system. Examples are measurements 
on household appliances or portable tools. 

CAT III:

 For measurements performed in the building installation at the 

distribution level such as on hardwired equipment in fixed 
installation and circuit breakers. 

CAT IV: 

For measurements performed at the primary electrical supply 
(<1000V) such as on primary overcurrent protection devices, 
ripple control units, or meters 

 

Receiving Your Shipment 

Both models include a 9V battery, warranty card and user manual. 
Upon receiving your shipment, be sure that the contents are consistent 
with the packing list. Notify your distributor of any missing items. If the 
equipment appears to be damaged, file a claim immediately with the 
carrier and notify your distributor with a detailed description of any 
damage. Save the damaged packing container to substantiate your claim. 

 

Description 

The MR probes are designed to meet the latest safety and performance 
standards. Two different hook-shaped jaws are offered, both permitting 
the user to “pry” into or “hook” onto cables (will accept 2 x 500 MCM) or 
even smaller bus bars. 

The MR series uses Hall effect technology. The electronics and batteries 
are self-contained in the handles. The output of the AC/DC probes is 
1mV/A and 10mV/A. An auto zero push button ensures rapid and stable 
zeroing. There is no output filtering - True RMS with DC components is 
possible. Phase shift is excellent, making the MR series well suited for 
power and power quality applications. 

Model MR411 is a portable 400A AC (600A peak), 600A DC current 
probe. Model MR521 is a portable 1000A AC (1500A peak), 1500A DC 
current probe that accurately measures AC or DC current waveforms 
using Hall effect technology. 

Both models have proportional mV output for direct readings on 
multimeters, recorders, loggers and other instruments accepting banana 
plugs. 

Содержание MR411

Страница 1: ...AL OL ON 600A OFF 60A 600V CAT III 600A OUTPUT 600A 1 mV A 60A 10 mV A I N S T R U M E N T S AC DC CURRENT PROBE Model MR 411 OL ON 1500A OFF 150A 600V CAT III 1500A OUTPUT 1500A 1 mV A 150A 10 mV A I...

Страница 2: ...me of purchase or obtained by returning the instrument to our repair and calibration facility for a nominal charge The recommended calibration interval for this instrument is 12 months and begins on t...

Страница 3: ...Model MR Probes 12 DC Measurement 12 AC Measurement 13 Indicator Lights 13 Auto Off 13 Operation Examples 14 DC Current Measurement Example MR411 14 Measuring the DC Component of AC DC Waveform MR411...

Страница 4: ...circuit or input ALWAYS connect the Model MR411 or MR521 to the display device before clamping the probe onto the sample under test ALWAYS inspect the instrument and lead prior to use Replace any def...

Страница 5: ...stributor with a detailed description of any damage Save the damaged packing container to substantiate your claim Description The MR probes are designed to meet the latest safety and performance stand...

Страница 6: ...ions ELECTRICAL Current Range 150A Range 0 2 to 100A AC 150A Peak 0 4 to 150A DC 1500A Range 0 5 to 1000A AC 1400A Peak 0 5 to 1500A DC Output Signal 10mV A on 150A range 1mV A on 1500A range Accuracy...

Страница 7: ...r two 0 95 24mm or two bus bars 1 24 x 0 39 31 5 x 10mm Dimensions 8 8 x 3 82 x 1 73 224 x 97 x 44mm Weight 15oz 440g MR521 Specifications cont Overload 1500A DC and 1000A AC continuous up to 1kHz Noi...

Страница 8: ...eading DC to 440Hz Battery 9V alkaline NEDA 1604A IEC 6LR61 recommended Low Battery Green LED when battery voltage 6 5V Battery Life Approx 100H typical Overload Indication Red LED indicates input gre...

Страница 9: ...increments of 25 to 40mA Red LED goes on at push of the button and turns off when zero is reached no need to the hold button down Case protection IP30 per IEC529 Drop Test 1 0m on 38mm of oak on conc...

Страница 10: ...00 4 2 Test voltages 4kV level 2 in contact aptitude criteria B 8kV level 3 in the air aptitude criteria B Radiated Field IEC 1000 4 3 1995 With max interference of 5 of the measurement range 3V m lev...

Страница 11: ...8 9 7 Green light on when battery voltage 6 5V 8 Red light overrange and incorrect zero adjustment 9 Protective non slip guard 1 Jaws 2 Conductor 3 Automatic zero DC button 4 Safety banana plug 4mm 5...

Страница 12: ...8 9 1 Jaws 2 Conductor 3 Automatic zero DC button 4 Safety banana plug 4mm 5 Lead 5ft 1 5m 6 Range selection switch On Off 1500A 1mV A 150A 10mV A 7 Green light on when battery voltage 6 5V 8 Red lig...

Страница 13: ...pable of displaying 1mV of output per amp of measured current Voltmeter accuracy of 0 3 or better to take full advantage of the probe accuracy Input impedance of 1M 100pF or greater When the probe is...

Страница 14: ...wer will automatically shut off see Auto Off on next page DC Measurement Select the appropriate DC mV or V range on the DMM or measuring instrument Zero the probe With the probe disconnected from test...

Страница 15: ...s Replace the 9V battery if the green LED is not lit The red LED indicates a momentary or continuous overload of the instrument Readings taken while the red LED is on or flashing should be considered...

Страница 16: ...Examples DC Current Measurement Example for Model MR411 Conductor carrying 25A DC in the direction of the arrow DMM placed in DC volts mode DMM displays 250 0mV with the probe in the 60A 10mV A range...

Страница 17: ...suring the DC Component of an AC DC Waveform for Model MR411 Conductor carrying 300A DC 25 0A AC DMM placed in DC volts mode DMM displays 300 0mV with the probe in the 600A 1mV A range Figure 4 ACV DC...

Страница 18: ...ing the AC Component of an AC DC Waveform for Model MR521 Conductor carrying 900A AC 20A DC Voltmeter placed in AC volts mode Voltmeter displays 900 0mV with probe in the 1500A 1mV A range Figure 5 AC...

Страница 19: ...it is important to select the range that provides the best resolution Make sure that probe jaw mating surfaces are free of dust and contamination Beware of short circuit currents Large in rush DC cur...

Страница 20: ...6 8 10 12 14 Primary current in amperes Typical Accuracy on the 60 A DC Range Typical Accuracy at 50 Hz Range 60 A 10 mV A Typical Accuracy at 50 Hz Range 600 A 1 mV A Error in 1 10 100 1000 2 0 2 4 6...

Страница 21: ...150 A Range in DC 1 1 5 1 0 5 0 0 5 1 1 5 2 10 100 1000 Primary current in amperes Accuracy in Typical Accuracy on the 1500 A Range in DC 1 1 0 1 2 3 4 10 100 1000 10 000 Primary current in amperes Ac...

Страница 22: ...ng Clean the body of the clamp with a cloth lightly moistened with soapy water Wipe clean with a cloth moistened with clean water and dry Do not use solvent Warranty The Models MR411 and MR521 are war...

Страница 23: ...S T includes calibration certificate plus recorded calibration data Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Tel 800 945 2362 Ext 360 603 749 6434 Ext 360 Fax 603...

Страница 24: ...Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 www aemc com 99 MAN 100093 v13 05 17...

Отзывы: