background image

NOTE:

 For more information on using the DataView

®

 software, refer to the Model 

6292 user manual that is supplied on the USB stick.

Getting Started

 

WARNING: 

Read the user manual and safety warnings before using this instrument. 

Safety procedures and rules for working near high voltage energized systems must 
be observed during the use of this instrument. The generated voltages and currents 
may be dangerous.

1. 

Connect the instrument to a 120/240V 50/60Hz AC power supply.

2.

  Perform a test with the test leads shorted. If the current does not reach the 

preset current (or the measured resistance is significantly higher than 0

W)

check the connection and try again. If the problem persists, contact technical 

assistance. 

3. 

Before turning the instrument on, connect the test probes to the device to be 

tested and to the appropriate front panel terminals.

Connection Example in BSG Mode:

 

WARNING: 

Do not connect or disconnect the test leads during a measurement.    

Grounding connections are strongly recommended to protect the operator and 
equipment. Use a strong grounding point as close as possible to the device under 
test. This ground is required to verify the instrument is working properly.  

Saving a Test

• 

To save a test result, press 

SAVE

 during the test to get partial results or at 

the end of the test to save the final result.

•  If the Auto Save function in the menu settings is set to ON, the final result 

will be automatically saved at each successfully completed test.

Содержание Model 6292

Страница 1: ...Micro Ohmmeter Model 6292 CHAUVIN ARNOUX GROUP www aemc com ENGLISH Quick Start Guide...

Страница 2: ...the instrument to our repair and calibration facility for a nominal charge The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer For...

Страница 3: ...th the instrument Micro Ohmmeter Model 6292 Cat 2129 83 Set of Two 25 ft Kelvin Clips 200A Hippo Cat 2129 72 Large Classic Tool Bag Cat 2133 71 Green Ground Lead w Clamp Cat 2129 88 USB Cable 5 ft Cat...

Страница 4: ...nt is protected by double insulation Battery Important instructions to read and to fully understand Useful information or tip to read The CE marking guarantees conformity with European directives and...

Страница 5: ...C terminal Ensure the terminals are free of any voltage in relation to earth ground point and between each other Take into account that in a substation you will find in disconnected points high potent...

Страница 6: ...ector 8 LCD Display 9 Fuse 10 Power cord connector 11 Fan 12 Alphanumeric keypad and function buttons 13 Rotary test knob 14 USB communication port 15 ON OFF switch Current Probe Connector 1 2 3 4 5 1...

Страница 7: ...e rotary knob has two primary purposes 1 adjust the current during a test and 2 navigate through menus and select options When the main screen is displayed a short press will perform the MENU button f...

Страница 8: ...the connection and try again If the problem persists contact technical assistance 3 Before turning the instrument on connect the test probes to the device to be tested and to the appropriate front pan...

Страница 9: ...nnection to the instrument Then from the menu bar select Instru ment then click Configure The Configuration Panel dialog box will appear This dialog box allows you to configure options for each of the...

Страница 10: ...you confirm that you want to proceed with the deletion all objects are immediately removed from memory irrespective of whether you save any other changes made through the Configuration Panel dialog b...

Страница 11: ...MA 02035 USA Phone 800 343 1391 508 698 2115 Fax 508 698 2118 E mail techsupport aemc com NOTE Do not ship instruments to our Foxborough MA address Limited Warranty The Model 6292 is warranted to the...

Страница 12: ...08 17 99 MAN 100394 v7 Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Phone 603 749 6434 Fax 603 742 2346 www aemc com...

Отзывы: