  Precautions for Use  

The protection obtained by the instrument can be compromised if it is used in a 

way that is not recommended by the manufacturer.

• 

Do not attempt to perform any tests with this instrument until you have 

read the user manual. 

• 

Tests are to be carried out on de-energized circuits only! Never connect 

the unit to a live circuit.  

• 

The micro-ohmmeter must be connected to the earth/ground point, through 

the ground terminal or the power cord.

• 

Be sure the power cord is accessible at all times, in case the instrument 

needs to be quickly unplugged for any reason. Always disconnect the 

power cord before performing any repair on the instrument.

• 

Do not obstruct either the intake vents or the fan, to avoid overheating.

• 

During a circuit breaker measurement its contacts must be closed and 

connected to an earth/ground point. The end connected to an earth/ground 

point must be connected to the “C-” terminal. 

• 

Ensure the terminals are free of any voltage in relation to earth/ground 

point and between each other. Take into account that in a substation 

you will find, in disconnected points, high potential levels in relation to 

the earth/ground point. Those potentials are caused by presence of 

electromagnetic fields and can be minimized following the indications in 

the paragraph before. 

• 

Make sure that the current connections are well connected, as well as the  

C clamp, to avoid undesirable heating. 

• 

Be careful when manipulating the current terminals in the instrument.  

High temperatures may occur in the current connections. 

• 

Never connect or disconnect the cables during a measurement. If a 

modification must be made, press the STOP button first.

•  The micro-ohmmeter should never be used in an explosive environment 

(this includes poorly ventilated battery rooms and enclosures).

• 

The instrument, test leads and measuring wires must be free of defects 

and should be changed if there is any evidence of deterioration (insulation 

split, burnt, etc.).

•  Never exceed the safety values indicated in the specifications.

Содержание Model 6292

Страница 1: ...Micro Ohmmeter Model 6292 CHAUVIN ARNOUX GROUP www aemc com ENGLISH Quick Start Guide...

Страница 2: ...the instrument to our repair and calibration facility for a nominal charge The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer For...

Страница 3: ...th the instrument Micro Ohmmeter Model 6292 Cat 2129 83 Set of Two 25 ft Kelvin Clips 200A Hippo Cat 2129 72 Large Classic Tool Bag Cat 2133 71 Green Ground Lead w Clamp Cat 2129 88 USB Cable 5 ft Cat...

Страница 4: ...nt is protected by double insulation Battery Important instructions to read and to fully understand Useful information or tip to read The CE marking guarantees conformity with European directives and...

Страница 5: ...C terminal Ensure the terminals are free of any voltage in relation to earth ground point and between each other Take into account that in a substation you will find in disconnected points high potent...

Страница 6: ...ector 8 LCD Display 9 Fuse 10 Power cord connector 11 Fan 12 Alphanumeric keypad and function buttons 13 Rotary test knob 14 USB communication port 15 ON OFF switch Current Probe Connector 1 2 3 4 5 1...

Страница 7: ...e rotary knob has two primary purposes 1 adjust the current during a test and 2 navigate through menus and select options When the main screen is displayed a short press will perform the MENU button f...

Страница 8: ...the connection and try again If the problem persists contact technical assistance 3 Before turning the instrument on connect the test probes to the device to be tested and to the appropriate front pan...

Страница 9: ...nnection to the instrument Then from the menu bar select Instru ment then click Configure The Configuration Panel dialog box will appear This dialog box allows you to configure options for each of the...

Страница 10: ...you confirm that you want to proceed with the deletion all objects are immediately removed from memory irrespective of whether you save any other changes made through the Configuration Panel dialog b...

Страница 11: ...MA 02035 USA Phone 800 343 1391 508 698 2115 Fax 508 698 2118 E mail techsupport aemc com NOTE Do not ship instruments to our Foxborough MA address Limited Warranty The Model 6292 is warranted to the...

Страница 12: ...08 17 99 MAN 100394 v7 Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Phone 603 749 6434 Fax 603 742 2346 www aemc com...

Отзывы: