10
Digital Thermometer Model K2000F and Temperature Probe Model ST2-2000
Surface Sensor SK5
Working Range:
-58°F to 932°F (-50°C to 500°C)
Time Constant:
1 second
Length:
5.9”
Class:
II (NF C 42-322 standard)
This sensor, for flat surfaces, is fitted with a spring loaded tip which per-
mits good contact, even if the sensor is not exactly perpendicular to the
surface. The use of silicon grease improves the quality of contact.
Supple Sensor SK6
Working Range:
-58°F to 545°F (-50°C to 285°C)
Time Constant:
1 second for contact measurement
3 seconds for ambient air measurement
Length:
39”
Class:
I (NF C 42-322 standard)
This supple sensor is both thin and long. It is recommended for measure-
ment areas that are difficult to reach. It must not be used with liquids
because the tip is not water resistant. It has a Kapton sheath (derived from
Teflon).
Air Sensor SK7
Working Range:
-58°F to 482°F (-50°C to 250°C)
Time Constant:
5 seconds
Length:
5.9”
Class:
I (NF C 42-322 standard)
Suitable for all ambient air measurement. The extremely small sensitive
element is protected by a metallic shroud. During measurement in still air
conditions, it is recommended that the probe should be waved from side
to side, or up and down if you prefer.