CHAPTER 5

MAINTENANCE

Use  only  factory  specified  replacement  parts.  AEMC

®

  will  not  be  held 

responsible  for  any  accident,  incident,  or  malfunction  following  a  repair 

done other than by its service center or by an approved repair center.

5.1 

Changing the Batteries

NOTE: 

It is recommended that all data is downloaded before replacing the 

batteries.

 

WARNING: 

Turn the unit off before changing the batteries or loss of recorded 

data may occur. Disconnect the unit from any voltage measurement points 
before opening the rear cover to change the batteries.

Remove the screw from the battery cover

• 

Slide off cover to remove

• 

Replace with only 1.5V AA alkaline batteries, then replace cover

• 

Press the 

• 

PRESS

 button for 2 seconds to turn ON

NOTE:

  If  the  instrument  is  stored  without  the  batteries  installed,  the 

internal clock will need to be reset using the Simple Logger

®

 II Control 

Panel software once the batteries are installed.
Only store the unit with the batteries installed for short periods of time. For 

prolonged storage of the unit, it is recommended to remove the batteries.

 

WARNING: 

Pressing the 

RESET

 button can result in the loss of recorded data. 

The instrument will try to recover normally, however under certain conditions, 
the clock may reset and the memory may fill. The clock may then need to 
be set and memory erased before another recording can start. 

The RESET 

button should only be pressed as a last resort.

An additional memory backup capacitor is used to provide backup power 

while the main batteries are being changed. This backup power will last for 

approximately 10 seconds while batteries are not installed.

www.

ShopAEMC

.com

Shop for AEMC products online at:

1.888.610.7664

Содержание CL601

Страница 1: ...E N G L I S H User Manual SIMPLE LOGGER II TRMS CLAMP ON MODEL CL601 www ShopAEMC com Shop for AEMC products online at 1 888 610 7664...

Страница 2: ...ed at the time of purchase or obtained by returning the instrument to our repair and calibration facility for a nominal charge The recommended calibration interval for this instrument is 12 months and...

Страница 3: ...fety 9 4 OPERATION 10 4 1 LED Control Operation and Status Function 10 4 2 Connecting the Simple Logger II to a Computer 12 4 3 Turning the Unit On 12 4 4 Recording Data 13 4 4 1 Starting a Recording...

Страница 4: ...MAINTENANCE 19 5 1 Changing the Batteries 19 APPENDIX A TROUBLESHOOTING 20 APPENDIX B GLOSSARY 21 Limited Warranty 23 Warranty Repairs 23 www ShopAEMC com Shop for AEMC products online at 1 888 610 76...

Страница 5: ...reinforced insulation Indicates aWARNING and that the operator must refer to the user manual for instructions before operating the instrument Failure to follow or carry out any instructions preceded b...

Страница 6: ...ents performed at the primary electrical supply 1000V such as on primary overcurrent protection devices ripple control units or meters 1 3 Receiving Your Shipment Upon receiving your shipment make sur...

Страница 7: ...e of up to eight times per second The measurement process is performed no more than eight times per second TRMS calculations are performed on a single line cycle This means that the input s are ignore...

Страница 8: ...Components and Features 1 Control Button PRESS This button marked PRESS selects the mode of operation Use this button to start or stop recordings erase the memory clear alarms and turn the instrument...

Страница 9: ...tor 4 Reset Button not shown The RESET button resets the CPU and is located in the battery com partment To access this button remove the battery compartment cover then reset the instrument by insertin...

Страница 10: ...ally isolated Communication Rate 115200 bps Power Source 2x1 5V AA cell alkaline batteries Battery Life 100 hours to 45 days dependent on sample rate and recording length This unique recording mode pr...

Страница 11: ...aw Opening 1 65 42mm max Vibration IEC 68 2 6 1 5mm 10 to 55Hz Shock IEC 68 2 27 30G Drop IEC 68 2 32 1m 3 3 Environmental Operating Temperature 14 to 122 F 10 to 50 C Storage Temperature 4 to 140 F 2...

Страница 12: ...the button will result in no action being taken This provides a mechanism to cancel or ignore the button press Control of the instrument is performed by pressing and holding the PRESS button while the...

Страница 13: ...he logger is not OFF Note Overload occurs when any input is 10 above its input range When the battery voltage goes below 1 7 volts the instrument will shut down ter minating the recording if it is rec...

Страница 14: ...LEDs light instantaneously the instrument was in SLEEP mode Releasing the PRESS button will show the status Connect the instrument to a USB port on your computer and establish communication with the...

Страница 15: ...NDBY mode see 4 3 3 Press and hold the PRESS button When the START GREEN LED lights up release the button 4 The GREEN LED double blinks when the logger is recording 4 4 2 Stopping a Recording Session...

Страница 16: ...mory using the PRESS Button 1 Press and hold the PRESS button When the ERASE RED LED lights up release the button This will arm the instrument for an erase operation when not in record mode While arme...

Страница 17: ...the storage rate recording period and measurement format using the Configure Instrument dialog box in the SLII Control Panel software Trend measurements are stored at this fixed storage rate 4 9 Norm...

Страница 18: ...the memory is partially full To save clear or check memory availability use the Simple Logger II Control Panel software There may be instances where the GREEN LED is also double blinking indicating t...

Страница 19: ...elow the usable level or if the batteries have been removed for an extended period of time the clock time will be lost However the recorded memory will be maintained since it is stored in Flash memory...

Страница 20: ...If the logger is not responding to a button press make sure the unit has sufficient battery power If the battery voltage is below 1 7V the unit will not respond to a button press In this condition pr...

Страница 21: ...N NOTE If the instrument is stored without the batteries installed the internal clock will need to be reset using the Simple Logger II Control Panel software once the batteries are installed Only stor...

Страница 22: ...Make sure the Simple Logger II is properly configured so that you have Storage Rate Recording Period and at least one Measurement Channel specified Symptom The instrument does not respond to a button...

Страница 23: ...ure of frequency equivalent to cycles per second I O Input output a device or port capable of sending or receiving digital information Port A name given to any connector allowing input or output of in...

Страница 24: ...www ShopAEMC com Shop for AEMC products online at 1 888 610 7664...

Страница 25: ...strument to us for repair provided we have your warranty registration information on file or a proof of purchase AEMC Instruments will at its option repair or replace the faulty material Warranty Repa...

Страница 26: ...24 Simple Logger II Model CL601 Notes www ShopAEMC com Shop for AEMC products online at 1 888 610 7664...

Страница 27: ...www ShopAEMC com Shop for AEMC products online at 1 888 610 7664...

Страница 28: ...05 08 99 MAN 100308 v4 Chauvin Arnoux Inc d b a AEMC Instruments www ShopAEMC com Shop for AEMC products online at 1 888 610 7664...

Отзывы: