18
Power Quality Analyzer Model 8335
CHAPTER 3
SPECIFICATIONS
3.1 Reference Conditions
Parameter
Reference Conditions
Ambient temperature
73°F ± 5°F (23°C ± 3°C)
Humidity (relative humidity)
[45%; 75%]
Atmospheric pressure
[860 hPa; 1060 hPa]
Phase-to-neutral voltage
[50Vrms; 1000Vrms] without DC (< 0.5%)
Standard current circuit input voltage
(except AmpFlex
®
& MiniFlex
®
)
[30mVrms; 1Vrms] without DC (< 0.5%)
*A
nom
⇔
1Vrms
3 × *A
nom
÷
100
ó
30mVrms
Rogowski current circuit input voltage
(Only AmpFlex
®
& MiniFlex
®
)
[11.73 mVrms; 117.3mVrms] without DC (< 0.5%)
3000 Arms
⇔
117.3mVrms to 50Hz
300 Arms
⇔
11.73mVrms to 50Hz
Frequency of electrical network
50/60Hz ± 0.1Hz
Phase shift
0° (active power and energy)
90° (reactive power and energy)
Harmonics
< 0.1%
Voltage unbalance
< 10%
Voltage ratio
1 (unity)
Current ratio
1 (unity)
Power supply
Battery only
Electric field
< 1 V.m-1
Magnetic field
< 40 A.m-1
NOTE:
The symbol “U” will be used throughout this manual and in the
instrument to refer to phase-to-phase voltage measurement.
The symbol “V” will be used for phase-to-neutral voltage measurement.
*The values of A
nom
are provided in the next table.