45
Specific Specifications in MAX/MIN mode (from 10Hz to 5kHz in voltage and
from 10Hz to 1kHz in current):
•
Accuracy: add 1% R to the values of the table above.
•
Capture of the extreme: approximately 100ms.
4.2.20 Specifications in THDr
Measurement Range
0.0 to 100%
Specified Measurement
Range
0 to 100%
of the measurement range
Accuracy
± (5% R ±2cts) in voltage
± (5% R ±5cts) in current
Resolution
0.1%
4.2.21 Specifications in THDf
Measurement Range
0.0 to 1000%
Specified Measurement
Range
0 to 100%
of the measurement range
Accuracy
± (5% R ±2cts) in voltage
± (5% R ±5cts) in current
Resolution
0.1%
Note
: The display is "----" if the input signal is too low (V<5V or I<6A) or if
the frequency is less than 5Hz.
Specific characteristics in MAX/MIN mode
in THD (from 10Hz to 1kHz):
•
Accuracy: add 1% R to the values in the tables above.
•
Capture of the extreme: approximately 100ms