34
5.1.8
REPEAT, DELETE, GO TO AND VIEW COMMANDS
·
Repeat command enables operator to repeat a specified number of wedges/test
points
·
Delete command enables operator to disregard (leave without any data) a specified
number of wedges/test points
·
Go To command enables operator to “jump” to any predefined test point and
continue the test.
·
View command provides quick view of test results for all wedges tested during the
test.
To delete or repeat measurement(s) the user should specify number of wedges/test points,
counting backwards from the last tested wedge.
Figure 31: Delete Measurement
Figure 32: Repeat Measurement
Figure 33: Go To specified location
NOTE: Only in Sequential Test, following Slots, instead of Number of wedges, dialog box
will ask for Number of Test Points to be deleted or repeated. In all other cases these
commands will be based on number of wedges, since software will automatically recalculate
number of test points per wedge. Furthermore, if more than one group is created, each with
different settings (following Slots or Wedges), the maximum number of measurements to be
specified using these commands will be limited to the number of test points in that group.
For example, if the operator wants to repeat the last ten measurements, having five of them
in Group 2, and five in Group 1, command Repeat should be used twice before re-testing:
first to return to the beginning of Group 2, and after that again to go backwards for five
wedges at the end of Group 1. Only the latest test data will be stored
.
Содержание WTD-501