R3131 Spectrum Analyzer Operation Manual
2.3 Measurement Examples
2-69
Jul 10/98
Setting the measurement conditions
This changes the analyzer settings so that the input signal is displayed more clearly.
5.
Press FREQ, 1, 8, 9, 5, ., 1, 5 and MHz.
A center frequency of 1895.15 MHz is set.
6.
Press SPAN, 2 and MHz.
A frequency span of 2 MHz is set.
7.
Press BW, RBW AUTO/MNL, 1, 0 and kHz.
The RBW is set to 10 kHz.
Setting the detector mode
8.
Press TRACE, Detector and Posi.
The trace is set to positive detector mode.
Setting the specified bandwidth
This sets the adjacent channel leakage mode and specifies the specified bandwidth.
9.
Press POWER MEASURE and ACP.
The ACP menu is displayed.
10.
Press Channel Band WD, 3, 0, 0 and kHz.
The specified band width is set to 300 kHz.
Performing ACP GRAPH
This measures the adjacent channel leakage power.
11.
Press Graph ON/OFF.
The fixed delta marker is displayed, and the operation result of the adjacent
channel leakage power measurement is displayed as a graph.