R3131 Spectrum Analyzer Operation Manual
2.2 Basic Operation
2-23
Apr 10/98
2.2.3
Measuring Window and the Display Line
This section describes the measuring window which is used to display measurements within a limited area,
and the display and reference lines which are used to compare traces.
Power on
1.
Turn the spectrum analyzer power on.
Initialization
This resets the current settings to the factory defaults.
2.
Press SHIFT and CONFIG (PRESET).
The default settings have now been reset.
Figure 2-12 Factory Defaults Screen
Calibration signal output
This activates the calibration signal used for the measurement.
3.
Press SHIFT, 7 (CAL) and Cal Sig Level ON/OFF.
Cal Sig Level ON/OFF is turned on, and the calibration signal is output.