ARMA-C125-CRM
37
The coupling loss may vary with frequency and the specific characteristics of the RF circuit.
Accordingly, a calibration chart is provided via the front panel display for each sample port. The
calibration chart for the port identifies the coupling loss at a number of frequencies across the
operating range of the amplifier. The loss associated with intermediate frequencies can be
obtained by interpolation.
7.10.1
M
ONITOR
C
ALIBRATION
C
HARTS
1.
To view the monitor calibration chart for the RF sample port on an amplifier, use the
[
é
é
]
and
[
ê
ê
]
keys on the front panel of the amplifier to access the Monitor Calibration sub-menu.
2.
Press
[Enter]
once to view the Output Monitor calibration chart followed by the input
monitor calibration chart. The display is capable of showing one line of the calibration chart
at a time. Use the
[
è
è
]
and
[
ç
ç
]
keys to scan the chart in order to obtain the coupling loss
associated with specific frequencies for each coupling port.
Example
: If the calibration factor CALFACT for the output monitor at 6.150 GHz = - 40.2 dB
and the external Power Meter measurement shows Pm = +6.8 dBm, then the SSPA output
power will be calculated as following:
P
out
= Pm + |C| = 6.8 dBm + 40.2 dB = 47 dBm.