27
Enter the desired shift start time, test interval and end time. Then, turn the validation reminder on. The
performance validation reminder is now active. When the current time matches the test interval the following
performance reminder screen will appear:
Selecting ‘Cancel’ simply returns the system to the main screen. Selecting ‘Begin’ navigates the user back to
the ‘Performance Validation Menu’ screen shown previously. Please note: the user will be prompted for the
level 3 access password when selecting ‘Begin’. From the ‘Performance Validation Menu’ screen select
‘Validation Ferrous’. The following screen will appear:
Test the product with the ferrous test wand. If a detection occurs press ‘Test Passed’. If it did not or an
unwanted detection occurs press ‘Test Failed’. The data logger will record the ferrous test wand size entered
previously and the date and time of the detection only if a detection occurs and the user presses ‘Test Passed’.
Otherwise no recording to the data logger is made noting the performance validation test. When the test is
complete the user will be sent back to the ‘Performance Validation Menu’. Select the ‘Validation Non-Ferrous’
Button and repeat the process. Repeat entire process for stainless.