Adtech UDB97 Скачать руководство пользователя страница 2

6.0  CALIBRATION 

To perform a calibration check or re-calibration of the instrument 

follow this procedure. 

A. Carefully remove the instrument from the housing to

gain  access to  the inside  components  observing  all

normal safety and equipment precautions.

B. Make sure the unit l/0 wiring is properly connected and

that  the  correct  power  source  per  the  label  is  also

connected.  The instrument must be at normal power

for a minimum of 2 minutes before proceeding to C.

C. Connect  the input contacts for  Up-Down and Ramp­

Hold to the appropriate inputs.

D. The output may be monitored either as a direct voltage

for a voltage output signal or as a current that can be

represented as a voltage across a resistor shunt.

E. Turn  the  multiturn  potentiometers  marked  UPRATE

and DOWN RATE fully CW.  This sets the Ramp rates

at the high value for t he particular jumper configuration.

F. Temporarily  set  jumpers  J14,  J15,  J16  and  J18  to

position 8 to set the output response at the fastest rate.

G. Set  the  input  contacts  to  drive  the  output  down  to

minimum value and adjust the multiturn potentiometer

marked ZERO to provide the minimum calibrated output

(e.g.) 4.00 ma± 0.01 ma de.

H. Set the input contacts to drive the output to maximum

value and adjust the  multiturn potentiometer marked

SPAN 

to provide the maximum calibrated output (e.g.)

20.00 ± 0.01 ma de.

I. Repeat steps G and H until within calibration;± 0.1 %.

J. From the RAMP TIME  SELECTION TABLE Section

8.0 look up the value of N corresponding to the smaller

of the UP RAMP TIME and DOWN RAMP TIME and

calculate the frequencies;

fup  =  4096N Hz, 

fdown  =  4096N

Tup 

Tdown Hz 

where T  and T d<lwn are UP RAMP TIME and DOWN 

RAMP TiME in seconds. 

K.

To  set  the  UP-RAMP  TIME:  connect  a  frequency

counter from pin  13 of 

zg 

to Term #5; set the input

contacts for UP-RAMP mode and adjust the multiturn

potentiometer marked UP-RATeto provide a frequency

equal to f up.

L. To set the DOWN-RAMP TIME:  connect a frequency

counter from pin 13 of 

zg 

to Term #5;  set the input

contacts  for  DOWN-RAMP  mode  and  adjust  the

multiturn  potentiometer  marked  DOWN-RATE  to

provide a frequency equal to f down.

M.

Repeat steps Kand L until within calibration;± 0.1 % of

frequency.

N.

Note  the  jumper  configuration  corresponding  to  the

value of N in the table and set jumpers J14, J15, J16

and J18  accordingly.  This  step provides the actual

desired UP 

RATE 

and 

DOWN RA 

TE.

0. This completes the calibration.

7.0  FIELD TROUBLE SHOOTING GUIDE 

This section offers a simple, first level trouble-shooting aid for 

an apparent instrument malfunction. 

SYMPTOM 

No output 

CORRECTIVE ACTION 

1. Check the input and output connections

carefully.

2. Check that the 

power 

supply polarity is

correct and that 

power 

is present on the

instrument terminals.

3. Check that the input contacts are correct

and  that  they  change  their  state  from

open to close.

4. If the output is a current signal (4-20 ma,

etc.),  make  sure  the  output  loop  is

complete and thatthe correct meter range

is selected.

All external checks are complete.  Problem 

seems to be internal. 

The following information is provided for a qualified technician 

or  serviceman  as  check  points  for  use  in  internal 

troubleshooting. 

CHECKPOINT/ 

COMPONENT 

(across) 

C20 

Term 1 (-) to Ei+) 
Term 1 (-) to E

1

(+) 

Term 1(-)to pin 19 of  Z

4

(+) 

VOLTAGE/RANGE 

26 ± 4 vdc 

12 ± 0.6 vdc 

-12 ± 0.6 vdc 

9.4± 0.5 vdc

-2-

190-A-000029-0

Отзывы: