7.2.4 Alarms
Here all the alarm signals that have been configured for this process or program are shown,
together with their current status.
Hint
Alarms are critical operating statu
s, which are sent to the„VisuNet“ recording program
and logged.
7.3 Alarm signal
If an alarm has occurred, for example through a sensor break or a digital input (alarm), the „Step“
box will be flashing alternately
(light and dark).
On the status bar the alarm and its cause
(e.g.
temperature sensor break)
will be shown.
Be aware
Temperature sensor break, program is terminated.
Core sensor break, program is terminated only when nominal value is pre-
programmed.
7.4 Program runtime, operating time display in operating mode
Press the “Time" button for about 3 seconds to display the program runtime and /or the
remaining program runtime while a program is in operation.
Current step time:
Runtime of a program
Program runtime:
Duration of the
current program
(goes beyond
individual steps.)
Alarm
Sensor break!
Remaining program runtime:
Time interval that remains until the end of
a program (goes beyond individual
steps). If a step without step time limit
exist in the remainder of the program
sequence, the value 00:00 will be shown.
With nominal step time = 00:00
Step runtime
With a nominal step time of >00:00
Remaining runtime of the step
Program runtime
3 Sec.
MKA 500
MKA 800
Remaining
program runtime
3 Sec.
MKA 500
MKA 800