The way to hold range is to make instrument automatically measure, and hold range after
measuring out correct and stable value, or select range according to the impedance of the
device under test, referring to the next table.
LCR range:
Range No.
Range
Measurement range
8
10Ω
0
̚
10Ω
7
30Ω
10Ω
̚
70Ω
6
100Ω
70Ω
̚
300Ω
5
300Ω
300Ω
̚
1kΩ
4
1kΩ
1kΩ
̚
3kΩ
3
3kΩ
3kΩ
̚
10kΩ
2
10kΩ
10kΩ
̚
30kΩ
1
30kΩ
30kΩ
̚
100kΩ
0
100kΩ
100kΩ
̚
∞
Press
RANGE
key, and the range changes between AUTO and HOLD. Press UP or DOWN
key to manually select range. And the current range No. is displayed in range area on LCD.
AD2817C also has DC resistance measurement function. Besides the foregoing range,
another range is extended with different measurement range.
DCR range:
Range No.
Range
Measurement range
9
3
Ω
0
̚
3
Ω
8
1
0Ω
3
Ω
̚
1
0Ω
7
30
Ω
10
Ω
̚
100
Ω
6
100Ω
100
Ω
̚
3
00Ω
5
300Ω
300Ω
̚
1kΩ
4
1kΩ
1kΩ
̚
3kΩ
3
3kΩ
3kΩ
̚
10kΩ
2
10kΩ
10kΩ
̚
30kΩ
1
30kΩ
30kΩ
̚
100kΩ
0
100kΩ
100kΩ
̚
2M
Ω
4.2.8 Open correction
Press SHIFT + OPEN to enter open correction menu.
AD2817C
’s open correction function can eliminate the effect of test cable,
test fixture and
stray admittance in parallel with the device under test, such as stray capacitance.
Perform the following steps to perform open correction:
1. Press
SHIFT
key at test status, and “
SHIFT
” at left corner is on.
2. Press
OPEN
key to select correction function, as Figure 4-2 shows on LCD.