AD2817C Operation Manual
52
8
9
NC
NC
No Connection
No Connection
10
11
/OUT
/AUX
Bin sorting judgments.
All bin sorting signal outputs are opto-isolated open
collector.
12
13
/TRIG
/TRIG
External Trigger:
AD2817A is triggered on the rising edge of a pulse applied
to this pin when the trigger mode is set to EXT mode.
14
15
EXV2
EXV2
External DC voltage 2:
DC voltage supply pins for DC isolated control signals
(/TRIG, /IDX, /EOM). Setting of internal jumpers must be
changed when using an internal voltage supply.
16
17
18
VCC
Internal voltage supply (maximum 100 mA)
Setting of internal jumpers must be changed when using an
internal voltage supply.
19
/PHI
Primary Parameter High Reject:
The measurement value is greater than the high limit value
of BIN1
–
BIN3.
20
/PLO
Primary Parameter Low Reject:
The measurement value is less than the low limit value of
BIN1-BIN3.
21
/SREJ
Secondary Parameter Reject:
The measurement value is not within the range of the
secondary parameter limit.
22
23
24
25
26
NC
NC
NC
NC
NC
No Connection
No Connection
No Connection
No Connection
No Connection
27
28
EXTV1
EXTV1
External DC voltage 1:
DC voltage supply pins for DC isolated open collector
outputs, /BIN1-/BIN3, /AUX, /OUT, /PHI, /PLO, /SREJ. The
setting of internal jumpers must be changed when using the
internal voltage supply.
29
NC
No Connection
30
/IDX
/IDX signal is asserted when an analog measurement is
completed and AD2817A is ready for the next DUT to be
connected to the UNKNOWN terminals. The measurement
data, however, is not valid until /EOM is asserted.
31
/EOM
End of Measurement:
This signal is asserted when the measurement data and
comparison results are valid.
32
COM2
Common for EXTV2