Replacing the rechargeable batteries

 

The battery compartment is located at the front of the unit. 

1)  Open the battery compartment. 

2)  Remove  the  battery  cable  and  the  corresponding  connector  carefully  from  the  housing.  Disconnect  the 

battery  pack.  Connect  the  wires/plug  (red  to  red,  black  to  black)  with  the  new  battery  pack  and  insert 
same into the housing. 

3)  Close the battery compartment. 

Only  use  original  rechargeable  batteries  and  charger  units  from  ADE.  The  warranty  does  not 
cover any damage caused by using other types of batteries or charger units. 

Tips for handling the batteries 

The life of a battery depends essentially on the load, frequency of use, temperatures, settings and the acces-
sories used. 

– 

New batteries or batteries that have been stored for long periods require a longer charging period. 

– 

Charge the batteries at room temperature. 

– 

Store the batteries empty (uncharged) in a cool, dark and dry place. 

– 

Never expose the batteries to temperatures below -10°C (14°F) or above 45°C (113°F). 

– 

Over the course of time, batteries lose their charging capacity and require a recharging period. If the bat-
tery performance drops considerably, then it is time to replace the battery. 

– 

New batteries are supplied only in partial charged condition. Some batteries only achieve their full capaci-
ty after several complete charge and discharge cycles. 

– 

When disposing of disused batteries, observe the valid legal and municipal regulations. 

Never dispose of batteries in fire as they can leak or explode!

  

 

Quality features 

ADE  products  correspond  to  the  applicable  EC  Directives  and  European  standards  to  manufacture 
and  distribute  world-wide  medical  devices  as  per  the  latest  technical  standards  ensuring  a  long  life 
time and continuous reliability. 

 

This  symbol  stands  for  compliance  of  the  European  Directive  covering  medical 
devices and declares the conformity as per the Medical Device Law.. 

 

Approved  scales  classes  III  respecting  the  standards  of  the  European  guideline 
covering approved scales are carrying this sign. 

 

This symbol means: "Consult operating instructions". To achieve accurate results, 
read the instructions for use carefully before use and follow all instructions provid-
ed therein. Keep the instructions for use for future reference. 

 

Products carrying this symbol meet the requirements as per: 

  The valid EC Directive for non automatic scales  

in the latest version. 

Using the scales for legal metrological applications 

The scales are used in medical facilities for diagnostic and medicinal purposes as long as they are calibrated. 
The scales must be recalibrated according to legal conditions and regulations. 
 
 
 
 

Содержание M403020

Страница 1: ...Operating Manual Class III Approved Scales Electronic Chair Scales M403020 M403020 210322 Rev006 UM e...

Страница 2: ...ting additional devices 6 Housing and connections 7 Modes of operation 7 Settings 7 Setting date and time 8 Operation 8 Positioning of patient 8 Weighing 8 Resetting the scales back to zero 8 Tare fun...

Страница 3: ...ee at a dry place Alignment of the scale 1 Place the chair scale on a plain dry and level surface before the weighing of patients 3 The air bubble in the spirit level should be right in the centre of...

Страница 4: ...are supplied only in partial charged condition Some batteries only achieve their full capaci ty after several complete charge and discharge cycles When disposing of disused batteries observe the valid...

Страница 5: ...for a long period to enter and save the tare value 11 BMI Button Enter to proceed with BMI mode and to enter as next step the height Press again to return to weight mode 12 Units Button Switch button...

Страница 6: ...Only use approved accessories and peripheral equipment Only operate the unit under permitted ambient conditions Remove the mains plug before cleaning the unit Do not operate the unit in hazardous or u...

Страница 7: ...ommands are being set in this mode To start setting mode hold the GROSS NET key pressed until TIME appears on the display Configuration mode set up mode The values for different parameters including t...

Страница 8: ...t position before patient stand up again Armrest and footrest are not to be used as helping device for getting up from the chair They are to be understood as supporting device for arms and legs only i...

Страница 9: ...shown corresponds to the gross weight minus the tare value If there is no tare value saved the display remains in gross mode The tare value flashes briefly then after a second the display changes to...

Страница 10: ...rflow indicator negative Negative weight indicated weight too low to be displayed 99999 Overflow indicator positive Positive weight indicated weight too high to be displayed 999999 AD LOW A D above ne...

Страница 11: ...materials or workmanship All movable parts e g batteries cables power units etc are excluded This warranty does not cover any normal external appearance of use and wear or any damage occurring as a re...

Страница 12: ...t Full duplex RS 232 38400 19200 9600 4800 2400 1200 bps Printer Port Output only RS 232 or active 20 mA current loop 9600 4800 2400 1200 bps Both Ports 7 or 8 data bits even odd or no parity Operator...

Страница 13: ...formity by the manufacturer This product has been manufactured in accordance with the harmonised European standards They con form to the regulations of the EC directives listed below NAWI Directive 20...

Страница 14: ...14 M403020 210322 Rev006 UM e...

Страница 15: ...M403020 210322 Rev006 UM e 15...

Страница 16: ...16 M403020 210322 Rev006 UM e Dealer...

Отзывы: