7351 Series Digital Multimeter Operation Manual
8.11 Error Message List
8-22
8.11 Error Message List
This section describes the error number and how to deal with error displayed on this instrument.
No.
Error display
Error description
[ ]: Response to the ERR? command
Solution
1
Err1
RAM read/write error
[RAM read/write failed]
Repair
2
Err2
Display RAM test error or Communication error
[Front panel does not respond]
Repair
3
Err3
Calibration data check SUM error
[Calibration memory lost]
Repair
4
Err4
ROM check SUM error
[ROM Chk SUM]
Repair
5
Err6
Parameter check SUM error
[Save/recall memory lost]
Repair
6
Err16
FLASH memory writing error
[FLASH memory error]
Repair
7
Err101
The function cannot be set.
[Inhibited function]
Solution1
8
Err140
RS-232 communication error
[RS-232 Communication Error]
Solution10
9
Err150
USB communication error
[Illegal packet received]
Solution2
10
Err200
Ratio test error of IR1 to IR2 in AD
[AD Ratio 1-2]
Repair
11
Err201
Ratio test error of IR2 to IR3 in AD
[AD Ratio 2-3]
Repair
12
Err205
AC
1 zero measurement test error in analog operation
[AC
1 Zero Check]
Repair
13
Err206
L
1 zero measurement test error in analog operation
[L
1 Zero Check]
Repair
14
Err207
L
10 zero measurement test error in analog operation
[L
10 Zero Check]
Repair
15
Err208
L
100 zero measurement test error in analog operation
[L
100 Zero Check]
Repair
16
Err209
DC
10 zero measurement test error in analog operation
[DC
10 Zero Check]
Repair
17
Err210
Communication error in Analog
[Communication error (Analog)]
Repair
18
Err215
1/100 ATT measurement test error in analog operation
[ATT 1/100 Check]
Repair