Error Message:

Description:                         Solution:

Error Displays and Troubleshooting

Err

Heartbeat rate

detection error.

The cuff was deflated

too rapidly.

Keep the deflation rate

at about 2-3 mmHg

per second.

EE

Press the “     ” button

to switch the unit off, or

the unit will

automatically switch

off after

60 seconds.

E3

Data Error.

Remove and reinsert

the batteries. If this

error keeps occurring,

return the device to

your local distributor

or service center. 

01

The cuff was inflated to

a pressure over 300

mmHg.

Open the valve im-

mediately to   rap-

idly exhaust the air

in the cuff.

A Pressure of over

15mmHg remaining

for more than

3 minutes is detected.

Содержание Diagnostix E-sphyg 7002

Страница 1: ...1 e sphygTM User s Manual DigitalAneroid...

Страница 2: ...ormance You ll quickly appreciate the results for you now own one of the finest sphygmomanometers that money can buy With proper care and maintenance your ADC blood pressure instrument is sure to prov...

Страница 3: ...d Maintenance 12 Disinfection and Sterilization 12 Specifications 13 Standards 13 Limited Warranty 14 Product Registration 15 This manual is intended to assist the user in the safe and efficient opera...

Страница 4: ...digital e sphygTM aneroid also has several features not available in a more traditional mechanical gauge The systolic assist function will give you a visual alert when you are ap proaching the patient...

Страница 5: ...igned for normal blood pressure measurement at the upper arm or thigh on healthy skin Proper use of this device is essential for accurate measurement Caution Do not inflate your digital aneroid unit t...

Страница 6: ...ometer consists of a digital manometer gauge complete inflation system ADCUFFTM calibrated nylon cuff latex free inflation bladder squeeze bulb and the ADFLOWTM valve a zippered carrying case and oper...

Страница 7: ...acing in the correct direction 4 Replace the battery cover and tighten the screw 5 Rotate the clip back Note You should replace the batteries if The weak battery mark appears in the display Nothing ap...

Страница 8: ...our hand during use ADCUFFTM nylon cuffs are specially designed to promote the accurate determination of blood pressure Index and range markings ensure use of the correct cuff size The artery mark ind...

Страница 9: ...display this information on the LCD screen once the cuff has been deflated Note Systolic assist and pulse display features are NOT motion tolerant Exces sive movement may interfere with accurate displ...

Страница 10: ...per second and stop to check at the following test points 299mmHg 250 mmHg 150 mmHg 100 mmHg and 50 mmHg Your gauge should be within 3 mmHg of your reference stan dard ADC recommends using the most se...

Страница 11: ...to switch the unit off or the unit will automatically switch off after 60 seconds E3 Data Error Remove and reinsert the batteries If this error keeps occurring return the device to your local distribu...

Страница 12: ...a damp cloth To help preserve it from cracking or discoloring wipe with a soft cloth moistened with ethanol Valves Clean the surface with a dry fabric such as soft cotton material Remove dust from fil...

Страница 13: ...ure 32 115 F 0 46 C Humidity 85 Relative Humidity max STORAGE ENVIRONMENT Temperature 4 131 F 20 55 C Humidity 90 Relative Humidity max non condensing Dimensions 3 L X 23 8 W X 11 8 H Weight 3 14 oz G...

Страница 14: ...by abuse mis use accident or negligence Incidental special or consequential damages Some states do not allow the exclusion or limitation of incidental special or consequential damages so this limitati...

Страница 15: ...00 232 2670 fax 631 273 9659 www adctoday com email info adctoday com 2015 ADC All rights reserved No one is permitted to reproduce or duplicate in any form this manual or any part thereof without the...

Отзывы: