Error Message:
Description: Solution:
Error Displays and Troubleshooting
Err
Heartbeat rate
detection error.
The cuff was deflated
too rapidly.
Keep the deflation rate
at about 2-3 mmHg
per second.
EE
Press the “ ” button
to switch the unit off, or
the unit will
automatically switch
off after
60 seconds.
E3
Data Error.
Remove and reinsert
the batteries. If this
error keeps occurring,
return the device to
your local distributor
or service center.
01
The cuff was inflated to
a pressure over 300
mmHg.
Open the valve im-
mediately to rap-
idly exhaust the air
in the cuff.
A Pressure of over
15mmHg remaining
for more than
3 minutes is detected.