Instructions for Use

Diagnostix

TM

Aneroids

Pocket, Palm, Clock Type

®

Inflate the cuff

Close the Adflow valve by turning thumbscrew clockwise.

Palpate the radial artery while inflating the cuff. Be sure to inflate cuff quickly by

squeezing bulb rapidly.

Inflate cuff 20-30 mmHg above the point at which the radial pulse disappears.

NOTE:

Cuff pressure range is 0 mmHg to 300 mmHg.

Position the Stethoscope

Position the chestpiece in the antecubital space below the cuff, distal to the brachium.

Do not place chestpiece underneath the cuff, as this impedes accurate measurement.

Use the bell side of a combination stethoscope for clearest detection of the low-

pitched Korotkoff (pulse) sounds.

Deflate the cuff

Open the valve to deflate the cuff gradually at a rate of 2-3 mmHg per second.

Measurement

Record the onset of Korotkoff sounds as the systolic pressure, and the disappearance

of these sounds as diastolic pressure. (Some healthcare professionals recommend

recording diastolic 1 and diastolic 2. Diastolic 1 occurs at phase 4).

NOTE:

It is recommended that K4 be used in children aged 3 to 12, and K5 should

be used for pregnant female patients unless sounds are audible with the cuff deflated,

in which case K4 should be used. K5 should be used for all other adult patients.

After measurement is completed, open valve fully to release any remaining air in the

cuff. Remove cuff.

Care and Maintenance

Pocket Gauge:

After measurement, fully exhaust cuff then wrap cuff around gauge

and bulb and store in zippered carrying case.

Palm Gauge:

After measurement, fully exhaust cuff and store in zippered carrying

case. If unit will not be used for an extended period of time, leave the trigger valve in

the full exhaust position.

Clock Gauge:

After measurement, fully exhaust cuff then wrap cuff around bulb and

store in storage compartment.

NOTE:

This product will maintain the safety and performance characteristics specified

at temperatures ranging from 50°F to 104°F (10°C to 40°C) at a relative humidity

level of 15% to 85%. This device can safely be stored at temperatures ranging from -

4°F (-20°C) to 131°F (55°C) with a relative humidity of 90%.

Manometer:

Your Diagnostix brand pocket aneroid gauge (#800 or #802), palm

aneroid (804N) or clock aneroid (805) requires minimal care and maintenance.

The manometer may be cleaned with a soft cloth but should not be dismantled under

any circumstances.

Should the indicator needle of the manometer rest outside

the oval calibration mark (Figure 5), then the manometer

must be re-calibrated to within ±3 mmHg when compared

to a reference device that has been certified to national or

international measurement standards. A manometer

whose indicator needle is resting outside of this mark, is

NOT acceptable for use.

In the event that the gauge is ever in need of calibration, simply return to ADC.

Damaged or broken parts will be replaced as needed at a minimum charge. Refer to

the warranty for specific details of warranty coverage.

The manufacturer recommends a calibration check every 2 years.

Cuff Cleaning and Disinfecting

Use one or more of the following methods and allow to air dry:

•    Wipe with mild detergent and water solution (1:9 solution).  Rinse.

•    Wipe with Enzol per manufacturer’s instructions.  Rinse.

•    Wipe with .5% bleach and water solution.  Rinse

•    Wipe with 70% isopropyl alcohol.

•    Launder with mild detergent in warm water, normal wash cycle.

      Remove bladder first. Cuff is compatible with 5 wash cycles.

Low-Level Disinfection

                                   

Prepare Enzol enzymatic detergent according to the manufacturer’s instructions. Spray

detergent  solution  liberally  onto  cuff  and  use  a  sterile  brush  to  agitate  the  detergent

solution  over  entire  cuff  surface  for  five  minutes.    Rinse  continuously  with  distilled

water for five minutes.  To disinfect, first follow the cleaning steps above, then spray

cuff with 10% bleach solution until saturated, agitate with a sterile brush over entire cuff

surface for five minutes.  Rinse continuously with distilled water for five minutes.  Wipe

off excess water with sterile cloth and allow cuff to air dry.

     CAUTION:

Do not iron cuff.

     CAUTION:

Do not heat or steam sterilize cuff.

Manometer Quality Control

A serial number and lot number are

automatically assigned to every aneroid 

during manufacturing, ensuring every 

item is controlled.

The serial number can be located on the 

faceplate of each aneroid (Figure 6).

The lot number is located on the outside

label of the manometer packaging

(Figure 7). 

Standards

    ANSI/AAMI/ISO 81060-1:2007, EN/ISO 81060-1:2012

Disposal

When your sphygmomanometer or any of its parts have reached their end of life,

please be sure to dispose of them in accordance with all regional and national 

environmental regulations. Devices that have become contaminated should be 

disposed of in accordance with all local ordinances and regulations.

(Figure 6)

Serial Number

Warranty

ADC

®

warrants its products against defects in materials and workmanship under 

normal use and service as follows: 

•  Warranty service extends to the original retail purchaser only and commences

    with the date of delivery.

•  Your Diagnostix™ manometer is warranted for life. The manometer is

    warranted to remain accurate to +/-3mmHg (or the prevailing standard) over
    its full range when compared to a reference standard for life.

•  Inflation system components (cuff, bladder, tubing, bulb, valves, and

    connectors) are warranted for three years.      

What Is Covered:

Replacement of parts, and labor.

What Is Not Covered:

Transportation charges to ADC. Damages caused by

abuse, misuse, accident, or negligence. Incidental, special, or consequential 

damages. Some states do not allow the exclusion or limitation of incidental, 

special, or consequential damages, so this limitation may not apply to you.

To Obtain Warranty Service:

Send item(s) postage paid to ADC, Attn: Repair

Dept., 55 Commerce Dr., Hauppauge, NY 11788. Please include your name and

address, phone no., proof of purchase, and a brief note explaining the problem.

Implied Warranty:

Any implied warranty shall be limited in duration to the terms

of this warranty and in no case beyond the original selling price (except where

prohibited by law). This warranty gives you specific legal rights and you may have

other rights which vary from state to state.

NOTE: 

Store gauge with

valve in full exhaust position.

(Figure 5)