7. Error Messages/Troubleshooting
If an error occurs during a measurement, the measurement is dis-
continued and a corresponding error code is displayed. We recom-
mend you switch the device off and take a new measurement making
sure you allow time for relaxation (Example: err 3).
Repeat measurement keeping arm
still.
If inflation of the cuff takes too long,
the cuff is not correctly seated or the
tube connection’s not tight.
Check connections and repeat.
The difference between systolic and
diastolic is excessive. Measure again
carefully following proper cuff proce-
dures and ensure measurement under
quiet conditions.
Error No. Possible cause(s) / Solutions
Err 1 The tube may have loosened, or no pulse
was detected.
*Ensure cuff connections are tight with
proper cuff placement. See Section 5.3
Err 2
Unnatural pressure impulses. Reason:
The arm was moved during the
measurement.
Repeat measurement keeping arm still.
Err 3
Err 5
The cuff pressure is too high. Relax
for 5 minutes and repeat the meas-
urement.*
HI
The pulse is too low (less than 40).
Repeat the measurement.*
LO
*If this or any other problem occurs repeatedly, please consult your
doctor.