Device Description and Intended Use
This device is intended to measure human body temperature in children and adults. It has a
compact size and digital display and can be used in a professional or household environment.
Parts and Assembly
Symbol Definitions
The following symbols are associated with your thermometer.
LIMITED WARRANTY
ADC
warrants its products against defects in materials and workmanship under normal houshold use and service as follows:
1. Warranty service extends to the original retail purchaser only and commences with the date of delivery.
2. Your Adtemp Digital Stick Thermometer is warranted for two years from date of purchase.
What Is Covered:
Replacement of parts, and labor.
What Is Not Covered:
Transportation charges to ADC. Damages caused by abuse, misuse, accident, or negligence.
Incidental, special, or consequential damages. Some states do not allow the exclusion or limitation of incidental, special, or
consequential damages, so this limitation may not apply to you.
To Obtain Warranty Service:
Send item(s) postage paid to ADC, Attn: Repair Dept., 55 Commerce Dr., Hauppauge, NY 11788.
Please include your name and address, phone no., proof of purchase, and a brief note explaining the problem.
Implied Warranty:
Any implied warranty shall be limited in duration to the terms of this warranty and in no case beyond the original
selling price (except where prohibited by law). This warranty gives you specific legal rights and you may have other rights which vary
from state to state.
For Australian Consumers:
Our goods come with guarantees that cannot be excluded under the Australian Consumer
Law. You are entitled to a replacement or refund for a major failure and compensation for any other reasonable foreseeable
loss or damage. You are also entitled to have the goods repaired or replaced if the goods fail to be of acceptable quality and
the failure does not amount to a major failure.
Visit
www.adctoday.com
to view this instruction booklet in other languages.
Low-battery Indication
Memory Indication
Measure Value
Measure Unit
Probe Tip
Flexible Probe Tip
Display Window
Multifunction Button
Symbol
Definition
Important warning/caution
Not made with natural rubber latex
Phthalate free
Operating instructions
Manufacturer’s information
Authorized European representative’s information
LATEX
Symbol
Definition
Meets essential requirements of European
Medical Device Directive 93/42/EEC
Lot number
Multifunction button
Auto detection error
Memory indication
Low-battery indication
M
0044
Specifications
Size
5.51”L x .78”W x 0.43”H (12.7 cm L x 1.9 cm W x .95 cm H)
Weight
0.63 oz (18 grams)
Display:
LCD, Liquid Crystal Display
Battery:
1.55V (LR41)
Battery Life:
Estimated 4,000 measurements
Measurement Time:
Oral 18-20 Seconds / Rectal 10 Seconds / Axillary 40 Seconds
Measurement Range:
90.0°F to 109.9°F / 32.0°C to 43.9°C (
±
0.2°F /
±
0.1°C)
Below 90.0°F (32.0°C) displays fixed “- - -” with fixed °F or °C.
Above 109.9°F (43.9°C) displays flashing “H” with fixed °F or °C
and 10 times “Beep” sound.
Accuracy:
0.2°F/0.1°C between 93.2°F to 107.6°F (34.0°C to 42.0°C)
a room temperature of 64.4°F to 82.4°F (18°C to 28°C)
±0.4°F/0.2°C between 90.0°F to 93.2°F (32.0°C to 34°C) and
107.6°F to 109.9°F (42.0°C to 43.9°C)
Operating Conditions:
50.0°F to 104.0°F (10.0°C to 40.0°C)
Storage Conditions:
-13°F to 131°F (-25°C to 55°C)
Relative Humidity:
15-95% Non-Condensing
Measurement Mode:
Real measurement with peak-hold method
Audible Alarm:
1 beep at start-up. Continuous beep at measurement completion
Auto Turn Off:
In approximately 8 minutes
Memory Feature:
Stores last temperature reading
Expected Service Life:
5 years
Standards:
Complies with EN 12470-3, ASTM E 1112-00, IEC 60601-1,
IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-56
Electromagnetic compatibility: device fulfills the
stipulations of
standard IEC 60601-1-2
The manufacturer recommends a calibration check every 2 years.
IB p/n 93-415FL-00 rev 1
Printed in China
LATEX
Onbo Electronic (Shenzhen) Co., Ltd.
No. 497, Ta Laneg Nan Road, Ta Laneg Street
Baoan District, Shenzhen, China
Microlife AG, Espenstrasse
139, 9443 Widnau, Switzerland
Dist. by: ADC
®
55 Commerce Drive, Hauppauge, NY 11788
Inspected in the USA
Made in China
tel: 631-273-9600
toll free: 1-800-232-2670
fax: 631-273-9659
www.adctoday.com
email: [email protected]
0044
EC
REP
Refer to user manual before use
Type BF applied part
Batteries and electronic devices must be
disposed of in accordance with the locally
applicable regulations, not with domestic waste.
93-415FL-00 rev 1.qxp_ADTemp 93-412-00 rev 3 CE 5/7/19 11:27 AM Page 1