0-Auto
Initial Zero range at power on.
(Options: 0%, 2%, 5%, 10% and 20%)
Sets the range for setting a zero when
power is applied. The total capacity of the
scale will not be affected by the new zero
if it is set within 10%. If it is set to a higher
value, the scale may zero but the range
may be limited by load cell, A/D converter
or mechanical stops within the base.
P1 10
(10%)
0-rAnG
Zero range when
[Zero]
is pressed.
(Options: 4%,10%, 20%, 50%,100%)
Sets the range for setting a zero when
[Zero]
is pressed. The total capacity of
the scale will not be affected by new zero
if it is set within 10%. If it is set to a higher
value, the scale may zero but the range
may be limited by load cell, A/D converter
or mechanical stops within the base.
P2 10
(10%)
SPEEd
Sets the display update speed.
(Options: 7.5, 15, 30 or 60).
SPd 15
The
P2 CAL
sub-parameter has the basic functions used by the scale to set
the capacity, decimal point position, increment size as well as the
calibration procedures. Do not change the weighing functions as the
scales may not perform as specified. Either the capacity will be wrong,
the decimal point will be in the wrong position or the scale will increment
with values that are not as specified for the scales.
dEC1
Selects the decimal point position
(Options: 0, 0.0, 0.00, 0.000).
C 0.00
InC
Selects the increment size.
(Options: 1, 2, 5 and 10).
inC 10
CAP
Sets the scale capacity. The maximum value
for the scale is as specified for the scale. For
example 60kg.
P2 CAL
CAL
Starts the calibration procedure. See the
details of calibration procedure in section 10.
The sub-parameter “
P3 Pro
” is used to view the calibration constant used
by the scale, view the A/D counts and set the scale back to default
@Adam Equipment Company 2007