13.0 PARAMETER SETTING 

 
This  section  allows  the  user  to  access  the  parameters  for  customising  the 
scale. The parameters are split into 2 groups-  

 

1. Check weighing Parameters  
2. Scale Parameters 
 
In the following sections use the [Limit] key to scroll through the options, the 
[Tare]  key  to  accept  the  option  and  the  [Zero]  key  to  return  to  weighing.  
Wherever  numeric  values  are  required  to  be  entered,  use  the  [Limit]  and 
[Unit] keys to increment the flashing digits as explained in section 4.1. 

 
 

13.1  CHECK WEIGHING PARAMETERS 

 

To  enter  this  section  press  and  hold  the  [Limit]  key  for  4 seconds. You will 
be  asked  for  the  password  if  the  current  check-weighing  password  is 
anything  other  than  “0000”.  See  the  parameter  “

F4  PS

”  in  this  section.  Enter 

the  correct  password  using  the  numeric  entry  method  as  mentioned  in 
section 4.1. If the password is “0000” it will go directly to “

F1   LLK

”.  

 

Parameter  Description 

Options 

Default setting 

F1  LLk 

Limit  Lock  will  prevent  a  normal 
user  from  changing  the  check 
weighing  limits.    To  change  the 
limits,  this  parameter  must  be 
disabled  by  setting  it  to  off  or  the 
user must enter the password.

 

on 
oFF

 

 

oFF

 

F2  LEd 

Setting the LED

 

on 
oFF

 

on 

 

F3  bEP 

This  parameter  sets  the  Beeper  to 
off  or  on.  If  set  to  on,  the  beeper 
can  further  be  set  to  sound  when 
the  weighing  result  is  within  or 
outside the check-weighing limits.

 

bP oFF

  - Off 

bP inL

   - Within 

limits 

bP otL

 - Outside 

limits (>20d)

 

bP inL

 

F4 PS 

This  parameter  allows  setting  of  a 
new Check weighing password, 
If the old parameter is “0000” enter 

the  new  password  twice  when  “

P1 

_ _ _

“ & “

P2 _ _ _ _

” are displayed. 

When  complete,  it  will  display 

To be entered 
manually. 
 

0000  

Содержание WBW 15aM

Страница 1: ... Adam Equipment Company 2011 Adam Equipment WBW SERIES WBW M EC and WBW aM NTEP Type Approved P N 303669031 Revision E3 November 2011 ...

Страница 2: ... Adam Equipment Company 2011 ...

Страница 3: ...OPERATIONS 11 10 1 ZEROING 11 10 2 TARING 11 10 3 WEIGHING 12 10 4 CHECK WEIGHING 12 10 5 LIMITS STORED IN MEMORY 13 11 0 RS 232 INTERFACE 13 12 0 CALIBRATION 14 12 1 WBW M APPROVED SCALES CALIBRATION PROCEDURE 14 12 2 SECURITY SEALS 14 12 3 CALIBRATION COUNTER 15 12 4 WBW SERIES NOT APPROVED 15 13 0 PARAMETER SETTING 17 13 1 CHECK WEIGHING PARAMETERS 17 13 2 SCALE PARAMETERS 18 14 0 ERROR CODES 1...

Страница 4: ... Adam Equipment Company 2011 2 ...

Страница 5: ...eries The WBW series are similar to the WBWa series except the WBW scales are usually set for metric units whereas for the WBWa series are usually set for imperial units However the user can change the setting as explained in section 13 2 All have stainless steel weighing platforms on an ABS plastic base assembly which is sealed to IP 65 making it water proof All scales have sealed keypads with co...

Страница 6: ...kg 16 000 kg Readability 0 0002 kg 0 0005 kg 0 001 kg 0 002 kg Repeatability S D 0 0002 kg 0 0005 kg 0 001 kg 0 002 kg Linearity 0 0004 kg 0 001 kg 0 002 kg 0 004 kg Grams Maximum Capacity 2000 g 4000 g 8000 g 16000 g Tare Range 2000 g 4000 g 8000 g 16000 g Readability 0 2 g 0 5 g 1 g 2 g Repeatability S D 0 2 g 0 5 g 1 g 2 g Linearity 0 4 g 1 g 2 g 4 g Pounds Maximum Capacity 5 lb 9 lb 18 lb 35 l...

Страница 7: ...acity tracker and symbols for units Scale Housing IP 65 Sealed ABS Plastic housing with Stainless Steel pan Pan Size 210 x 173 mm Overall Dimensions wxdxh 231 x 265 x 153 mm Net Weight 3 3 kg 7 26 lb Applications Weighing Scales Functions Weighing Check weighing WARNING The WBW M and WBW aM scales should not be calibrated by the user Calibration of the scales may make it illegal to use the scales ...

Страница 8: ...educe the accuracy Avoid extremes of temperature Do not place in direct sunlight or near air conditioning vents Avoid unsuitable tables The table or floor must be rigid and not vibrate Avoid unstable power sources Do not use near large users of electricity such as welding equipment or large motors Do not place near vibrating machinery Avoid air movement such as from fans or opening doors Do not pl...

Страница 9: ...le in the spirit level is in the centre of the level and the scale is supported by all four feet Attach the power supply module to the bottom of the scale and plug into the mains Press the O I key to start The scale will first display the software revision followed by a self test At the end of the self test it will display zero weight if the zero condition has been achieved and the weighing unit t...

Страница 10: ...he results This is the net weight Enters the selected parameter or value for setting Unit Selects the weighing units from a preset list of available units Shifts the flashing digit to the next digit when entering a value Limit Sets the limits for check weighing and allows setting of either the low limit or the high limit or both Increments the flashing digit or moves to the next option during sett...

Страница 11: ...he units Capacity Tracker A bar graph indicating the proportion of the scale capacity being used by the weight on the pan bAt LO Low battery Indicates full battery strength Will show less number of bars for weaker strength HI OK LO The scale is in Check weighing mode The colons are used to separate pounds from ounces Above the LCD to the left there are three LED s that indicates when the weight is...

Страница 12: ...2 hours for full capacity Below the display is a LED to indicate the status of battery charging When the scale is plugged into the mains power the internal battery will be charged If the LED is green the battery is being charged If it is red it is nearly discharged and yellow indicates the battery is increasing the charge level Continue to charge overnight for a complete re charge 8 0 BACKLIGHT Th...

Страница 13: ...ro to re zero the scale if small amount of weight is still shown when the platform is empty 10 2 TARING Zero the scale by pressing Zero The zero indicator will be on Place a container on the pan and its weight will be displayed Press Tare when the reading is stable The weight that was displayed is stored as the tare value and it is subtracted from the display leaving zero on the display The stable...

Страница 14: ...er The display will show the weight and the unit of weight currently in use To change the weighing unit press the Unit key The available weighing units are the ones that are enabled by the user in the parameters section See section 13 2 10 4 CHECK WEIGHING Check weighing is a procedure to show a display or cause a beeper to sound when the weight on the scale meets or exceeds the values stored in m...

Страница 15: ...gh limit or enter the new high limit using the numeric entry method see section 4 1 When the desired value is entered press Tare to accept the value The LO symbol will be on Display will show the last used low limit Set the low limit in the same way the high limit was set Pressing the Tare key will return the scale to weighing with the Check weighing function enabled NOTE The limits are displayed ...

Страница 16: ...d The scale has got metallic loops on the base and the cover located at the right rear corner as shown here A metal pin or wire can be passed through the loops and sealed using a lead wire seal Ensure the seal is secured enough to prevent removing the cover without breaking the seal or damaging the enclosure An alternative method would be to use tamper evident labels over the screw on the bottom c...

Страница 17: ...he WBW non approved scales are calibrated using metric weights when the weighing unit selected is either kilograms or grams and using pound masses when the weighing unit selected is either pounds ounces or pounds ounces Method To start the calibration turn the scale off and then turn it on again Press Tare during the self test Scale will show P Enter code number 0000 using the numeric entry method...

Страница 18: ...e to the mass and then return to weighing Remove the calibration weight as soon as calibration is complete NOTE If an error message FAIL H or FAIL L is shown during calibration re check the calibration and repeat if necessary If the error cannot be corrected contact your dealer or Adam Equipment for advice ...

Страница 19: ...rameter F4 PS in this section Enter the correct password using the numeric entry method as mentioned in section 4 1 If the password is 0000 it will go directly to F1 LLK Parameter Description Options Default setting F1 LLk Limit Lock will prevent a normal user from changing the check weighing limits To change the limits this parameter must be disabled by setting it to off or the user must enter th...

Страница 20: ...on of the scale Parameter Description Options Default setting S1 Un Enable or disable weighing units will not allow to disable all units at least one has to be enabled kg g lb oz lb oz kg S2 bL Backlight set to always on always off or automatic on whenever a weight is placed or a key is pressed EL oFF EL on EL AU EL AU S3 AoF Auto Off Disable or set time increment to turn off scale SLP 0 SLP 1 SLP...

Страница 21: ...he scale on Excessive weight on the pan when zeroing the scale Platform is not installed Improper calibration of the scale Damaged load cell Damaged Electronics Err 6 A D count is not correct when turning the scale on Load cell is damaged Electronics is damaged Err 8 High limit input error Low limit is set first then the high limit is set lower than the low limit and high limit not equal to zero E...

Страница 22: ... ready A Details of your company Name of your company Contact person s name Contact telephone e mail fax or any other methods B Details of the unit purchased This part of information should always be available for any future correspondence We suggest you to fill in this form as soon as the unit is received and keep a print out in your record for ready reference Model name of the scale WBW_____ Ser...

Страница 23: ...inal packaging and with correct documentation for a claim to be processed All claims are at the sole discretion of Adam Equipment This warranty does not cover equipment where defects or poor performance is due to misuse accidental damage exposure to radioactive or corrosive materials negligence faulty installation unauthorised modifications or attempted repair or failure to observe the requirement...

Страница 24: ...lichkeit 2004 108 EC Compatibilité électromagnétique This declaration is only valid when accompanied by a Certificate of Conformity issued by a Notified Body Diese Erklärung gilt nur in Verbindung mit einer Konformitätsbescheinigung einer benannten Stelle Cette déclaration est seulement valide quand elle est accompagnée par une Attestation de Conformité délivrée par un Organisme Notifié Fabrikant ...

Страница 25: ... Adam Equipment Company 2011 23 APPENDIX ...

Страница 26: ... Adam Equipment Company 2011 ...

Страница 27: ...in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense Shielded interconnect cables must be employed with this equipment to insure compliance with the pertinent RF emission limits governing this device Changes or modifications not expressly approved by Adam Equipment could void the user s authority to ope...

Страница 28: ...Medical Scales Retail Scales for Price computing For a complete listing of all Adam products visit our website at www adamequipment com Copyright by Adam Equipment Co Ltd All rights reserved No part of this publication may be reprinted or translated in any form or by any means without the prior permission of Adam Equipment Adam Equipment reserves the right to make changes to the technology feature...

Отзывы: