8.0 CALIBRATION
The PGL series can only be calibrated with an external mass.
8.1
MANUAL CALIBRATION
Pressing the
[Cal]
key will start calibration. Calibration can also be called for after a
set time period as determined by the user (see section 12.5).
Calibration using External Calibration mass:
1)
Pressing
[Esc]
will abort the calibration at any time
2)
Check the display is at zero. Tare if necessary
3)
Press the
[Cal]
key
4)
The display will show the balance setting a new Zero condition by showing
“LOAD 0”.
Make sure the pan is empty then press the
[Enter]
key to continue
5)
The balance will then show the value of the calibration mass required
sounding a beep, for example
“LOAD 1000 g”
6)
If a different mass is desired press the up or down arrow key to show the
alternate masses possible for the scale.
PGL
203
PGL
303
PGL
2002
PGL
3002
PGL
4001
PGL
6001
PGL
8001
PGL
10001
PGL
12001
PGL
15001
PGL
20001
1
100 g
100 g
1 kg
1 kg
2 kg
3 kg
4 kg
5 kg
5 kg
5 kg
10 kg
2
200 g
200 g
2 kg
2 kg
3 kg
4 kg
5 kg
10 kg
10 kg
10 kg
20 kg
3
300 g
3 kg
4 kg
5 kg
6 kg
12 kg
15 kg
4
6 kg
7 kg
5
8 kg