Adam Equipment 

 
 

 
 

MDW – 250L 

DIGITAL PHYSICIAN SCALE 

 
 

 

Software EMW-250L-20080601 

 
                
 
 
 
 
 
 
 
 
 
 
 

ADAM EQUIPMENT CO. LTD.       

p.n. 4302, Rev. C4, November 2009 

 

Содержание MDW - 250L

Страница 1: ... Adam Equipment Company 2009 Adam Equipment MDW 250L DIGITAL PHYSICIAN SCALE Software EMW 250L 20080601 ADAM EQUIPMENT CO LTD p n 4302 Rev C4 November 2009 ...

Страница 2: ... Adam Equipment Company 2009 Page 1 Rev C3 June 2008 ...

Страница 3: ... INTRODUCTION 4 2 0 TECHNICAL SPECIFICATIONS 4 3 0 DISPLAY AND KEY DESCRIPTIONS 5 4 0 SETTING UP THE SCALE 6 5 0 FUNCTION 5 5 1 WEIGHING 5 5 2 MEASURING HEIGHT 6 6 0 CALIBRATION 6 7 0 USER PARAMETERS 8 8 0 RS232 COMMUNICATIONS 9 9 0 ERROR MESSAGES 13 10 0 WARNING 13 ...

Страница 4: ... Adam Equipment Company 2009 Page 3 Rev C3 June 2008 ...

Страница 5: ...SPECIFICATIONS Maximum capacity 250 kg 550lb Minimum capacity 2 kg 4lb Scale division 0 1Kg 0 2lb Height range 80 cm 210 cm Division of measurement 0 5 cm Display LED display Size of platform 375 X 275 mm Overall dimension 535 X 275 X 1040 mm Deadweight 13 kg Environment for Use Temperature 5ºC 40ºC Humidity 85 RH Power 12vAC 500mA adapter Battery Internal re chargeable 6V 4Ah 5 hours approx Calib...

Страница 6: ...weighing in Kilograms Lb Indicates when the scale is weighing in Pounds Hold Indicates when the scale has held the weight reading shown on the display It will flash until it locks into the stable reading when it will then remain on constantly when it has held the displayed reading When the battery is low this symbol will be light up Connect the adapter to recharge the battery CH The charge light w...

Страница 7: ...ews number 3 on the diagram to mount the pillar to the indicator Connector the plug and socket number 6 and 7 on the diagram and insert the pole into the pole holder Secure the pole to the pole holder by inserting the 2 screws number 9 on the diagram Secure the height rod into the height rod holder number 4 and 8 on the diagram 8 height rod holder 4 height rod 2 cable to indicator 1 indicator 10 p...

Страница 8: ...ng the weighing unit to kg or lb The LED will indicate the chosen weighing unit Overload display When FULL appears on the display it shows that the load on the platform is over the maximum capacity Under these circumstances it is necessary to reduce the load otherwise the sensor or the platform will be damaged Hold Function To lock the weighing result press the Print Hold key The LED will flash un...

Страница 9: ... 1 When in normal weighing mode with the scale at zero press and hold down TARE and ON OFF keys to enter the calibration mode 2 If the calibration switch is in the off position on the main PCB inside the scale the indicator will show CAL OFF and then exit this mode If the indicator shows CAL the scale is ready for calibration showing the calibration switch is in the on position If the calibration ...

Страница 10: ...t that you selected and press the TARE key to confirm the chosen standard weight that was selected earlier The displayed data will flash on the display and if the scale accepts the calibration data it will calculate and store the information into the EEPROM If an error has occurred the scale will display CAL Er and return back to step 4 for re calibration If the loaded weight is not within the ran...

Страница 11: ...OFF key to exit the set up mode 3 Parameters setting summary Parameter x xy Factory Set Setting A o t 00 15 05 Auto off time No auto off 00 01 15 minutes auto off time P H 0 1 2 1 0 Only Print Function 1 Only Hold Function 2 both HOLD and PRINT function pressed down less 3s this key works as Print function pressed down more than 3s this key works as HOLD function H t 0 4 0 Hold work time 0 no time...

Страница 12: ...7 7 7 7 NC 8 8 8 8 NC 9 9 9 9 Note The indicator pin1 4 6 7 8 and 9 are not connected The RS232 function will only operate if PH has been set to 0 or 2 8 2 When Parameter S F in section 7 is set to 0 No RS232 function It will not transmit or receive any data although the scale is equipped with RS232 The RS232 function can be only activated when scale is in normal weighing mode RS 232 output of wei...

Страница 13: ...uate for use as a time out value by remote controlling device 8 5 1 The length of the weight field will be 7 digit weight data one for minus sign one for decimal point two for measure unit e g lb kg Units of measure abbreviations are always lower case If the weight is overcapacity the scale will return nine characters the field of minus sign decimal point weight data is filled by If the weight is ...

Страница 14: ...1H2H3 CR ETX 3 Command Z CR 5ah 0dh Response Zero function is activated and it returns to current scale status just like pressing ZERO ON OFF button LF H1H2H3 CR ETX If ZERO function cannot be activated it will return to current scale status 4 Command T CR 54h 0dh Response TARE function is activated and then returns scale status just like pressing TARE button LF H1H2H3 CR ETX If TARE function cann...

Страница 15: ...ble 1 under capacity 0 not at zero point 0 not over capacity 1 1 at zero point 1 over capacity 01 normal work mode 10 hold work mode 00 not define 11 not define 0 gross weight 2 always 0 always 0 1 net weight 0 eeprom OK 3 1 eeprom error always 0 always 0 4 always 1 always 1 always 1 5 always 1 always 1 always 1 6 always 0 always 1 always 0 7 parity Parity parity ...

Страница 16: ...e without following the necessary instructions Do not jump while standing on the platform This may damage the sensor inside Do not move the weighing machine violently and abruptly It is recommended to move and put down the weighing machine gently It is suggested to wipe the stains with soft cloth soaked with detergent and to wipe later with soft cloth too Do not use organic solutions and boiled wa...

Страница 17: ... Adam Equipment Company 2009 Page 16 Rev C3 June 2008 ...

Страница 18: ... Adam Equipment Company 2009 Page 17 Rev C3 June 2008 ...

Страница 19: ...is equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense Shielded interconnect cables must be employed with this equipment to insure compliance with the pertinent RF emission limits governing this device Changes or modifications not expressly approved by Adam Equipment could void the user s aut...

Страница 20: ...Crane scales Medical Scales Retail Scales for Price computing For a complete listing of all Adam products visit our website at www adamequipment com Copyright by Adam Equipment Co Ltd All rights reserved No part of this publication may be reprinted or translated in any form or by any means without the prior permission of Adam Equipment Adam Equipment reserves the right to make changes to the techn...