EN
© Adam Equipment Company 2014
14
9596revD-Jul14
Calibration selection values for all model:
•
Press
[Zero]
key
,
the display will show
“Load”
then add the selected
calibration mass value to the pan.
•
After the calibration mass is added, wait for the stable symbol to come
on and then press
[Zero]
key, the display will show “
PASS”
, and then
return to normal weighing after 3 seconds.
•
If it is outside of the previous calibration value by more than 5% ,it will
show
“FAIL L”
or
“FAIL H”
indicating the new calibration weight was too
low(FAIL L) or too high (FAIL H) in respect to the previous calibration
value.
8.0 ERROR
MESSAGE
---ADO
A/D counts are too high
---ADL
A/D counts are too low
--ERR-
Wrong operation
Unit-E
All the units are disabled
Low battery indication
LBK 3/6a
LBK 6/12a
LBK 12/25a
LBK 30/65a
1Kg/2kg/3kg
2Kg/4kg/6kg
4Kg/8kg/12kg
10Kg/20kg/30kg
2lb/4lb/6lb
4lb/8lb/12lb
10lb/20lb/25lb
20lb/40lb/60lb
Содержание LBK- 6/12a