@Adam Equipment Company 2004
5
2.5 AUTO SLEEP
The crane scales have automatic sleep function. You can disable this
function, see section 4.0. When the scale weighs a sample below the
minimum capacity (20d) for more than the set time, it will automatically
enter the sleep mode. Only the last point will turn on at LED. Once you
press any key or add more weight on the hook, it will return to normal
weighing mode again. Please pull out the antenna before operating the
remote controller.
2.6 REMOTE CONTROLLER
The crane scales are supplied with remote control. You will find 4 keys
on the remote controller.
Key A: use as
[ZERO]
key
Key B: use as
[CHANGE]
key
Key C: use as
[LOCK]
key
3.0 CALIBRATION
Turn off the power if you are using the scale.
Then turn on the power, press
[ZERO]
and
[CHANGE]
key at the
same time during self-checking.
Display shows “
”, asking for entering the password.
Input password, use
[LOCK]
key to move the active digit,
[CHANGE]
key to increment a digit. Press the
[ZERO]
key to
confirm the password. The default password is “
”.
Display will show “
”, press
[CHANGE]
key until display shows
“
”.
Press
[ZERO]
key to enter calibration, display shows “
”.
Empty the hook, after
STABLE
indicator on, press
[ZERO]
key to
confirm, after few seconds, zero calibration is now complete.
After zero calibration, display shows a value of mass for span
calibration. You can change this weight by using the
[LOCK]
key
to move the active digit and
[CHANGE]
key to increment a digit.
After entering the value of the calibration mass, press the
[ZERO]
key to confirm. When display shows “
”, add the calibrate mass
on the hook after stable indicator on. Press the
[ZERO]
key again,
wait for few seconds, span calibration is now complete.
Содержание CRW Series