Adam Equipment Company 2011
5
4
REMOVING
THE
COVER
If it becomes required to remove the cover, you should be very careful
as pulling or pressing on the load cell during this procedure can damage
it very easily.
When removing the screws (4 small screws below the scale) hold the
cover to the base and ensure not to press on the balance case as the
load cells can be damaged. The load cell is attached to the base.
The load cells are protected by overload and under-load stops. These
stops are either plates that contact the cover or screws that are
accessible from the outside of the scales.
Refer to the Trouble-shooting Guide for the possible causes of any
problem.
5
LOAD
CELL
DAMAGE
5.1
5.1
A/D
COUNTS
Press
[Unit]
first and then press
[On/Off]
, while keeping the [Unit] key
pressed to enter the dealer parameters. The display will show the
version number and then request passwords.
Press
[Unit],[Unit], [Tare], [Unit], [Unit]
in sequence to enter the setting
mode.
Press the
[Unit]
key several times until you enter the calibration settings.
Then you can see the A/D count.
To see the every division A/D counts ,for example: the CQT202 with the
readability 0.01g.
First record the current A/D count, e.g 43860, then put on the 1g weight
to see the new A/D Count ,eg:44198. So the division A/D count is
(44198-43860)/100=3.38. The counts should bigger than 3.