EN
© Adam Equipment Company 2017
10
5.0 CALIBRATION
OIML TYPE APPROVAL:
A jumper is placed on the PCB of CBK-M models to prevent User Calibration
and the scale is sealed. If the seal is broken or tampered with, the scale needs to be re-verified by an
authorised certification body and re-sealed, before it is used legally. Contact your local metrology standards
office for further assistance.
5.1
CBK CALIBRATION
The CBK scales are calibrated using kilogram weights and CBKa scales are calibrated using pounds.
To start the calibration, either get into the calibration section through the Scale Settings (
“FUnC 3”
- see
Section 4.2) or turn the scale off and switch on again and then press [Tare] during the self-test. Enter code
number 0000 and press [Tare]. This will take you directly to the calibration section. The display will show
"UnLoAd"
Remove all weight from the pan and then press the [Tare] key when the scale is stable. After the Zero point
is set, the display will show
“Ld xx”
. Place the suggested calibration mass on the pan. It is best to use a
weight close to the full capacity of the scale. If the mass is different from the displayed value, enter the value
of the mass in whole numbers. The kg or the lb symbol will be on to show the active unit.
Press the [Tare] key when the stable indicator is on.
The scale will calibrate to the mass. When complete, it will display
“PASS”
and then either display
“S8 CAL”
(if
entered the calibration section through the Scale Settings as per section4.2) or return to normal weighing (if
entered directly). Remove the calibration mass.
If an error message
“FAIL H”
or
“FAIL L”
is shown, re-check the calibration and repeat. If the error cannot be
corrected contact your supplier.
6.0 BATTERY
1) The scales can be operated from the battery, if desired. The battery life is approximately 90 hours.
2) When the battery needs charging a symbol on the display will turn on. The scale will still operate for
about 10 hours after which it will automatically switch off to protect the battery.
3) To charge the battery, simply plug into the mains power supply. The scale does not need to be turned
on.
4) The battery should be charged for 12 hours for full capacity.
Содержание CBK 100a