background image

EN 

 
© Adam Equipment Company 2017 

10 

 

 

5.0  CALIBRATION 

 

OIML TYPE APPROVAL: 

A jumper is placed on the PCB of CBK-M models to prevent User Calibration 

and  the  scale  is  sealed.  If  the  seal  is  broken  or  tampered  with,  the  scale  needs  to  be  re-verified  by  an 
authorised certification body and re-sealed, before it is used legally. Contact your local metrology standards 
office for further assistance. 

 

5.1 

CBK CALIBRATION 

 
The CBK scales are calibrated using kilogram weights and CBKa scales are calibrated using pounds. 
 
To  start  the  calibration,  either  get  into  the  calibration  section  through  the  Scale  Settings  (

“FUnC  3”

-  see 

Section 4.2) or turn the scale off and switch on again and then press [Tare] during the self-test.  Enter code 
number  0000  and  press  [Tare].  This  will  take  you  directly  to  the  calibration  section.  The  display  will  show

 

"UnLoAd" 

 

 
Remove all weight from the pan and then press the [Tare] key when the scale is stable. After the Zero point 
is set, the display  will  show

  “Ld    xx”

.  Place  the  suggested  calibration  mass  on  the  pan.  It  is  best  to  use  a 

weight close to the full capacity of the scale.  If the mass is different from the displayed value, enter the value 
of the mass in whole numbers. The kg or the lb symbol will be on to show the active unit. 
 
Press the [Tare] key when the stable indicator is on.   
 
The scale will calibrate to the mass. When complete, it will display

  “PASS” 

and then either display 

“S8    CAL”

 (if 

entered the calibration section through the Scale Settings as per section4.2) or return to normal weighing (if 
entered directly). Remove the calibration mass. 
 
If an error message

  “FAIL  H”

 or

  “FAIL  L”

 is shown, re-check the calibration and repeat. If the error cannot be 

corrected contact your supplier. 

6.0  BATTERY  

 

1)  The scales can be operated from the battery, if desired. The battery life is approximately 90 hours. 
2)  When the battery needs charging a symbol on the display will turn on. The scale will still operate for 

about 10 hours after which it will automatically switch off to protect the battery. 

3)  To charge the battery, simply plug into the mains power supply. The scale does not need to be turned 

on. 

4)  The battery should be charged for 12 hours for full capacity. 

Содержание CBK 100a

Страница 1: ...and environmentally friendly packaging where possible As part of this initiative we have developed a short form manual that uses less paper and ink to describe the main functions of your new Adam sca...

Страница 2: ...Adam Equipment Company 2017...

Страница 3: ...THE DISPLAY 2 2 3 TARING 2 2 4 WEIGHING 3 2 5 PARTS COUNTING 3 2 6 CHECK COUNTING 4 2 7 PERCENT WEIGHING 5 2 8 ACCUMULATED TOTALS 6 3 0 RS 232 INTERFACE 7 4 0 PARAMETERS 8 4 1 CHECK WEIGHING PARAMETER...

Страница 4: ...symbol indicators are also displayed 2 2 ZEROING THE DISPLAY You can press the Zero key at any time to set the zero point When the zero point is obtained the display will show the indicator for zero...

Страница 5: ...a sample weighed If a container is to be used place this container on the platform before entering parts counting and press Tare Press Count to enter the Parts Counting mode The display will show the...

Страница 6: ...t Pressing the Tare key to enter the value will return the scale to weighing with the Check weighing function enabled 4 During parts counting and percent weighing the limits are set in the same way as...

Страница 7: ...nt weighing function is entered pressing Tare again will return to normal weighing Remove the sample weight Then any other weight placed on the platform will be displayed as a percentage of the origin...

Страница 8: ...ghts have been added This can continue for up to 99 entries until the capacity of display is exceeded To view the total in memory press the Print key when the scale is at zero The display will show th...

Страница 9: ...face The scale when connected to a printer or computer through the RS 232 interface outputs the weight with the selected weighing unit Specifications Connector RS 232 output of weighing data ASCII cod...

Страница 10: ...allows the user to enter the Percent weighing Function See Section 10 7 None Enabled always F2 LLk This parameter prevents the normal user from changing the limits with the help of a Limit Lock With...

Страница 11: ...to disable all units at least one has to be enabled kg g lb oz lb oz kg S2 bL Backlight set to always on always off or automatic on whenever a weight is placed or a key is pressed EL oFF EL on EL AU E...

Страница 12: ...suggested calibration mass on the pan It is best to use a weight close to the full capacity of the scale If the mass is different from the displayed value enter the value of the mass in whole numbers...

Страница 13: ...0005 lb 0 001 lb 0 002 lb 0 005 lb Linearity 0 0004 lb 0 0004 lb 0 001 lb 0 002 lb 0 004 lb 0 01 lb Ounces Maximum Capacity 128 oz 256 oz 256 oz 560 oz 1120 oz 1600 oz Readability 0 005 oz 0 005 oz 0...

Страница 14: ...able battery 90 hours operation Calibration Automatic External Display 6 digits LCD digital display with capacity tracker and symbols for units Scale Housing ABS Plastic Stainless Steel platform Pan S...

Страница 15: ...C available Battery Internal rechargeable battery 90 hours operation Calibration Not permitted Display 6 digits LCD digital display with capacity tracker and symbols for units Scale Housing ABS Plasti...

Страница 16: ...weight on the pan when zeroing the scale Platform is not installed Improper calibration of the scale Damaged load cell Damaged Electronics Err 6 A D count is not correct when turning the scale on Load...

Страница 17: ...her methods B Details of the unit purchased This part of information should always be available for any future correspondence We suggest you to fill in this form as soon as the unit is received and ke...

Страница 18: ...nce faulty installation unauthorised modifications or attempted repair or failure to observe the requirements and recommendations as given in this User Manual This product may include a rechargeable b...

Страница 19: ...mercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio...

Страница 20: ...Phone 49 0 4340 40300 0 Fax 49 0 4340 40300 20 e mail vertrieb aeadam de Adam Equipment S A Pty Ltd 7 Megawatt Road Spartan EXT 22 Kempton Park Johannesburg Republic of South Africa Phone 27 0 11 974...

Отзывы: