@Adam Equipment Company 2007
9
7.0 USER
CALIBRATION
ACTION DISPLAYS
Press the
[Tare/Zero]
key during the
self-test at power on. The scale will
ask to enter the password.
“ Pi n”
Default Password is 0000.
Enter “0” four times.
Password can be changed in technical
parameters.
Press
[Print]
.
“ Pi n” “ ----”
Select the scale to be set up by using
the
[Local/Remote]
key.
Press the
[Print]
key to enter the
technical section.
“tEch” “ LocAL” “ “
“tEch” “rEmo tE” “ “
Use the
[U.Wt.]
to select the weighing
unit to be used for setting of the
calibration. The arrow in the “
Weight
”
window will indicate the unit selected.
Press the
[Print]
key to continue.
“tEch” “ Uni t” “ “
You will enter the first parameter -
Calibration. Display will ask to unload
any weight on the platform.
Press the
[Print]
key to continue.
“Un L o Ad”
Enter the calibration weight to be
loaded and press
[Print]
.
Load the calibration weight onto the
scale and press
[Print]
again.
The Calibration is complete, the scale
will run the self-test during which the
weight should be removed.
“SEL” “ 0010”
“Lo Ad”