4.0 ERROR
CODES
ERROR CODE
DESCRIPTION
POSSIBLE CAUSES
Err 4
Initial Zero is greater than
allowed (4% of maximum
capacity) when power is turned
on or when the
[Zero/Enter]
key
is pressed.
Weight on the pan when turning
the scale on.
Excessive weight on the pan when
zeroing the scale.
Improper calibration of the scale.
Damaged load cell.
Damaged Electronics.
Err 5
Keyboard error.
Improper operation of the scale.
Err 6
A/D count is not correct when
turning the scale on.
Platform is not installed.
Load cell is damaged.
Electronics is damaged.
Err 7
Scale is not stable when setting
the percentage weighing.
The scale is not stable.
Improper operation of the scale.
FAIL H
or
FAIL L
Calibration error.
Improper calibration (should be
within
+10% of the factory
calibration). The old calibration
data will be retained until the
calibration process is complete.
Err 9
Scale is unstable.
There is vibration or draft making
the scale unstable.
Electronics may be damaged.