© Adam Equipment Company 2007
12
The parameters available are:
“F1”
“
CAL”
To
enter
the
Calibration
“F2”
“
rES”
Resolution
selection
“F3”
“
CAPA”
Select
capacity
“F4”
“
init” “ZEro”
Initial Zero Range
“F5”
“
rE” “ Zero”
Re-Zero range
“F6”
“
SCSivE” “ tArE”
Successive Tare Enable
“F7”
“
Ad” “CoUnt”
Display the A/D counts
“F8”
“
Zero”
Select
the
zero
mode
“F9”
“
Pin”
To set the password
“F10”
“
Lvd” “mode”
Low voltage detection
6.2.1
F1
-CALIBRATION
To enter the calibration parameter, press the
[Tare]
key when “
F1 CAL
” is displayed.
The display will instruct you to remove any weight from the scale, “
UnloAd
”. Press
[Tare]
.
NOTE:
USA models are calibrated using “lb” unit.
The display will tell you to add the first weight to the scale: “
Ld
1 OO
kg
”
Place the first calibration weight as suggested in the Table below. Wait for stability then
press the
[Tare]
key.
The display will tell you to add the second weight to the scale: “
Ld
2 00
kg
”
Place the second calibration weight as suggested in the Table below. Wait for stability then
press the
[Tare]
key.
The display will show “
PASS
” if the calibration is OK. Or it will show “
FAIL L
” if calibration
could not be completed correctly.
Remove the calibration weight. The scale will return to weighing.