2
Adam Equipment Company 2014
3.0 CALIBRATION
To calibrate the balance follow the instructions below.
I.
Turn on the scale and zero it.
II.
Press and hold
【
MODE
】
and
【→0←】
keys together for about 5 seconds
until the display shows internal codes value.
III.
Press
【
MODE
】
key once, the display will show CALL
→
[xxxx], (where xxxx is
the full capacity weight value).
IV.
Load calibration weight as per the displayed value, and the scale will be
automatically calibrated. After calibration, the scale will return to normal
weighing mode. Calibration is now complete.
4.0 MODEL
SPECIFICATIONS
Model
CB 501
CB 1001
CB 3000
Maximum
Capacity
500g
1000g
3000g
Readability
0.1g
0.1g
1g
Units of Measure
grams (default), lbs and oz if the mode key is
enabled
Operating
Temperature
0 - 40 degrees C
Power Supply
2 x AA batteries or 6v DC 200ma Adaptor
Calibration Mass
500g
1000g
3000g
Display
5 x 12mm/0.47 inches high digits
Pan Size
130mm/5.1 inches diameter
Overall
Dimensions
(w x d x h)
160 mm x 210mm x 65mm
6.3 in x 8.1 in x 2.6 in
Net Weight
385 g / 0.85 lbs