© Adam Equipment Company 2007
4
Date/Time
Real Time Clock (RTC), To print date and time information
(Dates in year/month/day, day/month/year or
month/day/year formats- Battery backed)
Display
3 x 6 digits backlit LCD displays
Housing
ABS Plastic with Stainless Steel Pan
Overall Dimensions
315 x 355 x 110 mm
315 x 355 x 535 mm
Net Weight
4.3 kg
4.7 kg
*NOTE:
AZ
extra
-P models have the Pillar Display.
2.0 TROUBLESHOOTING
1.
Unit does not
turn on
Check whether the scale is plugged into the power supply properly.
Check the operation of the charging circuit.
Check whether the battery is charged- LED should turn to green
from red.
Check the adapter output.
.
2. The
scale
weighs but is
unstable
Air drafts or vibration or unstable table.
Pan rubbing against case or not installed correctly.
Scale not installed properly.
Improper connections on ADC circuit.
3. Scale
shows
wrong weight
Scale not installed correctly.
Check the Calibration.
To check whether a part of product weighed is trapped between
the pan and the scale.
Incorrect adjustment of the mechanical stops.
4.
An error “
E4
”
displays on the
screen
Load cell damage.
Scale not installed properly.
Incorrect adjustment of the mechanical stops.
Factory calibration has been tampered with- check the ADC counts
and re-calibrate.
Содержание AZextra 15
Страница 2: ...Adam Equipment Company 2007...
Страница 20: ...7 0 WIRING DIAGRAM Adam Equipment Company 2007 18...
Страница 21: ...8 0 MECHANICAL ASSEMBLY Adam Equipment Company 2007 19...
Страница 23: ...9 0 PILLAR ASSEMBLY Adam Equipment Company 2007 21...
Страница 24: ...10 0 FRONT DISPLAY ASSEMBLY Adam Equipment Company 2007 22...
Страница 25: ...11 0 REAR DISPLAY ASSEMBLY Adam Equipment Company 2007 23...
Страница 26: ...12 0 PILLAR DISPLAY ASSEMBLY Adam Equipment Company 2007 24...