8.0 ERROR MESSAGES
When power is first applied to the scale a check is made to assure the A/D
counts are within a normal range. If an error is shown at this time the A/D
counts are either too high or too low.
EEEEE
A/D counts at turn on are too high.
ADLOW
A/D counts at turn on are below zero.
In either case, check the A/D counts as described above. Verify the load cell,
load cell wiring and the A/D counts are acceptable. Try to calibrate the scale
if the A/D counts are acceptable.
During normal operation, the display may show an error under these
conditions:
EEEEHI
Overload
EEEELO A/D value too low
Low battery, replace battery if used. Check the
external power supply if batteries are not used.
9.0 TROUBLE SHOOTING GUIDE
PROBLEM
POSSIBLE CAUSES
Display is blank
Mains power is turned off
No turn on test
Power supply not plugged in
Power supply faulty
Display is disconnected or faulty
After turn on test
Pan not installed
Error message is
Unstable weight
Displayed
Load cell damaged
Display is unstable
Drafts or air currents
Obstruction under pan
Sample is moving (animal weighing)
Vibrations through table
Temperature changed dramatically
Power supply faulty
Weight value incorrect
Calibration error, Recalibrate
Unit calibrated with inaccurate weight
Balance not level
Obstruction between sample and cover or
breeze shield
Wrong unit of weight displayed