@Adam Equipment Company 2004
9
Calibration
•
Wait for the stable indicator to be displayed,
•
Press
[Unit]
- the display will show
"
",
with the
" "
flashing.
Place the 200g weight on the platform. A beep will be heard
after a few seconds and the display will show
"
".
•
Press
[On/Off]
, to turn off the power. Press
[On/Off]
to turn on
the power again.
•
Place maximum weight (200g) on the platform to verify weighing
is correct. If not, repeat the calibration.
10.2 AUTO CALIBRATION FOR ALL OTHER MODELS
•
Before entering calibration mode, press
[On/Off]
to turn the
power on for 1 minute.
Enter calibration mode
•
Press
[On/Off]
to turn the power off.
•
Press
[Unit]
first and then press
[On/Off]
, while keeping the
[Unit]
key pressed. The display will show the A/D value.
Calibration
•
Wait for the stable indicator to be displayed,
•
Press
[Unit]
-the display will show
"
".
After 2 seconds,
the display will return to A/D value.
•
Place the maximum weight (eg.2000g for AQT2000) on the
platform.
•
After the stable indicator is displayed press
[Unit]
•
The display will show
"
".
The calibration in complete now.
•
Press
[On/Off]
to turn the power off. Press
[On/Off]
to turn on
the power again,
•
Place full weight (eg.2000g for AQT2000) on the platform to
verify weighing is correct. If not, repeat the calibration procedure.
Содержание AQT Series
Страница 11: ... Adam Equipment Company 2004 10 ...