4. Density measurement 

 

The balance is capable of measuring the density of either a solid or a liquid.    The balance 

can be used with the Adam Equipment Density Determination Kit as described below or by 

using the weigh below abilities of the balance. 

 

 

 

4.1  Density Measurement of Solids 

To determine the density of a solid it is necessary to weigh the sample while in the air and 

then to weigh it while it is suspended in a liquid of known density.  The balance will guide you 

through the steps of this procedure using messages on the display.   The type of liquid that 

can be used are set by the user.  It includes distilled water, alcohol and any other liquid as 

long as the user knows its density.   

 

  WATER (distilled water)  

  ALCOHOL (spirit 100% +/- 0.1%  in temperature 20

0

C), 

  OTHER (a user supplied liquid with known density)  

 

For water and alcohol you will be required to give the temperature, therefore the thermometer 

is included in the kit. 
 
Density checking relays on weighing a sample in the air  and then in the liquid.  
Result of the test is shown on the  display automatically when the test if finished.  
Result of measurement can be sent over the RS 232 interface to a printer or PC by   

pressing the Print key. 

   

 

 

Содержание AEA Series

Страница 1: ...Adam Equipment Company 2000 Adam Equipment DENSITY DETERMINATION KIT for AEA and AEP BALANCES OPERATOR S MANUAL ADAM EQUIPMENT CO LTD pn Rev A May 2002...

Страница 2: ...Adam Equipment Company 2000...

Страница 3: ...all the calculations are done by the balance Speed of test the result is shown immediately after the sample is placed into the water or the float is placed into the liquid Accuracy before every measu...

Страница 4: ...he contents of the kits are slightly different depending upon the series of balances used The kit consists of the following components 1 Glass Float of known volume with 2 string 3 hook 4 Thermometer...

Страница 5: ...ger When the capacity tracker shows some weight is on the pan the balance should begin operating again and show a value on the display Use only enough weight on the hanger wires to enable the balance...

Страница 6: ...the steps of this procedure using messages on the display The type of liquid that can be used are set by the user It includes distilled water alcohol and any other liquid as long as the user knows its...

Страница 7: ...u for M6 Density Select the option M6 01 to choose Solid Density Select the liquid to be used set the temperature for the water or alcohol or set the density for another liquid With F enter submenu M6...

Страница 8: ...At any time press Print enter to enter the Density mode Make sure the Density Kit is assembled correctly and there are no air bubbles on any of the parts in the liquid Tare the balance if necessary t...

Страница 9: ...Make sure there are no bubbles on the sample of or any other part in the liquid 4 After the result is stable the unit of mass is on press You will see results of the density measurement and other mess...

Страница 10: ...of the Density Determination Kit at the beginning of this section This float has a hanger attached with the precise volume of the flat stamped upon it Before proceeding with the density test enter the...

Страница 11: ...to buzz or Enter the new value for the volume Save choice Procedure Measurement of liquid density compares the weight of the glass float when measured in the air and in the liquid If necessary instal...

Страница 12: ...nt through the RS 232 interface to a printer or PC by pressing To begin enter the Density menu M6 as above and select the M6 01 Liquid Density Option Set the cursor next to M6 04 Run With F enter mode...

Страница 13: ...loat in the liquid and when result are stable press Enter On the display you will see the computed value of density for the liquid You will also see additional information W1 float mass in the air W2...

Страница 14: ...o be ON 2 Delete previous statistics data 3 Enter the Density mode 4 Collect the data by pressing the Print key 5 After the data is collected look at the results of statistics and print them 6 Return...

Страница 15: ...o clear the data by pressing the Enter key Enable the Statistics function if necessary by setting cursor next to parameter 01 Statistic Enter parameter 01 Statistic Start to buzz or Change parameter d...

Страница 16: ...lue to the balance memory for statistic The value of N will increase by one RECALLING After collecting some data the results can be viewed and printed Enter the Density mode submenu Set the cursor nex...

Страница 17: ...mple of the results is shown N 10 SUM 44 60 g cm3 X 4 45 g cm3 MIN 4 43 g cm3 MAX 4 48 g cm3 D 0 00 g cm3 SDV 0 00 g cm3 RDV 0 00 Return to the statistic submenu for the density mode Return to the Den...

Страница 18: ...ompletely covered by about 10mm of the liquid Be certain any air bubbles are removed Dry the sample completely between tests Be very careful not to transfer water to the top pan when using tweezers Sa...

Страница 19: ...13 0036728 0 1 0012932 0 3 p C o t cm g d t temperature of air o C p pressure of air The procedure would be to measure the mass of the float in air value A while it is submerged in distilled water val...

Страница 20: ...0 998 0 9985 0 999 0 9995 1 0 5 10 15 20 25 30 Series1 Density of Distilled Water versus temperature C 0 1 2 3 4 5 6 7 8 9 0 0 99984 0 99990 0 99994 0 99996 0 99997 0 99996 0 99994 0 99990 0 99985 0 9...

Страница 21: ...Adam Equipment Company 2000 21...

Страница 22: ...Adam Equipment Company 2000 22...