INSTALLATION
The AE 402 indicator is supplied with drivers to control external relays.
The drivers could be used to control a number of different relays
depending upon the users need. The relay drivers are isolated outputs
requiring the use of an external power supply for the relays.
The Relay Accessory is a product to allow quick and convenient
connections to the AE402 for control of external equipment or
indicators.
Connections to the AE042 are made at terminal strip P1:
Within the AE402 the output to drive external relays is on the circuit
board inside the enclosure. To gain access you must remove the 6
screws securing the front to the rear of the case. Pass the wires for the
relays through the grommet on the rear panel. The wires will connect
to the PCB using the terminal strip P1. See the AE402 User manual for
more details.
Any of the 4 relays can be connected to any of the 4 outputs. The
+12Vdc from the relays connects to the Common terminal.