-  25  -

8.  Functions  of  “Hold”  and  the  Backlight

  Press  “HOLD”  to  temporarily  pause  the  HPM-300Aoperation.  The  operation  will  be  suspended  until  “HOLD” 
is  pressed  again.  Pressthe

button  to  operate  the  LCD  backlight.

9.  Characteristics  of  the  Clamp  Sensor  (Electric  Current  Sensor) 

An  electric  current  sensor  can  be  used  for  measuring  electric  power  consumption  without  blocking  or 
connecting  a  circuit.  This  sensor  is  equipped  with  a  fireproof  and  electric-shock-free  case  for  hand 
protection  in  compliance  with  the  safety  standards.
An  electric  current  sensor  must  be  separately  purchased. 

 

①  Procedure  for  Use
    1.  Connect  the  +  terminal  of  the  electric  current  sensor  to  HPM-300A’s  Hi  Current,  and  the  –  terminal 
        to  the  15A  terminal.
    2.  When  the  electric  current  measurement  cable  is  positioned  exactly  in  the  middle  of  a  sensor,  the  most 
        accurate  measurement  value  can  be  obtained.
    3.  The  value  flashed  on  the  screen  is  the  value  of  the  RMS  current  running  through  the  cable 
        (1A  =  1mV).
②  Caution
    Measurement  cannot  be  done  when  more  than  two  cables  are  connected.
③  Safe  Usage
    1.  Do  not  use  a  probe  with  a  more-than-AC-600V  cable.
    2.  This  equipment  enables  the  simultaneous  measurement  of  a  high  electric  current  in  a  high-pressure 
        cable  up  to  1000A  (using  an  electric  current  sensor).  The  inappropriate  operation  of  the  equipment  can   
        cause  casualties.
    3.  Using  the  equipment  with  a  cable  with  damaged  coating  is  very  dangerous,  and  careless  contact  with
        the  equipment  can  cause  electric  shock.  Make  sure  to  wear  rubber  boots  when  using  the  equipment,     
        and  use  it  on  a  rubber  floor  with  verified  insulated  products.  Normal  operation  in  a  grounded  state 
        must  also  be  maintained  to  ensure  safety.
    4.  Do  not  open  or  repair  the  equipment’s  case  while  operating  the  equipment.
    5.  Model
          CA-1000:  100(W)  x  244(H)  x  40(D)mm  490g,
                            Range-200A,  1000A,
                            AC/DC  use.

www.biall.com.pl

Содержание HPM-300A

Страница 1: ...1 Consumption Standby Power Meter Manual Product Name HPM 300A AD Power Co Ltd www biall com pl...

Страница 2: ...urement 4 High Current Measurement Using Clamp Sensor 5 Wh Watt Hour or Energy Usage Measurement 6 Range Setup 7 PC Communication Method Supported PC program 8 Functions of Hold and the Backlight 9 Ch...

Страница 3: ...t with 0 5mW to 7 5 KW power consumption in a Plug Play way with clamp sensor it can measure up to 220 KW power consumption thus offering simple and easy operation even by users at production plants a...

Страница 4: ...it with full attention to ensure its safe operation Do not load general batteries only rechargeable batteries and do not use an AC adaptor while loading general batteries the Li ion batteries are buil...

Страница 5: ...8 mm H x 47 mm D Weight 600 g inclusive of the batteries Fuse 250V 15A Measurement Accuracy Voltage measurement accuracy 1 AC 1V 500V 0 3 rdg 3dgt 2 DC 1V 1 000V 0 3 rdg 3dgt Current measurement accur...

Страница 6: ...NERGY 6 Frequency indicator it displays the measured frequency 7 It displays power factor CO2 emissions electricity charge unit 8 It displays the measured reactive power unit 9 It displays the measure...

Страница 7: ...version of voltage and electric current measurement unit manual range CT Five time consecutive pressing will lead to conversion to the CT mode from a conversion mode of an electric current unit ENERGY...

Страница 8: ...formation refer to the instructions for use Meets EU s safety standard Caution Electric shock FCC Meets U S s safety standard Initial Default Condition When the power is on the initial condition is AC...

Страница 9: ...9 Measurement Methods www biall com pl...

Страница 10: ...using a clamp sensor extended measurement is possible 2 2 220kW Caution To prevent an accident during the measurement make sure that the location of HPM 300A is in an appropriate mode before connecti...

Страница 11: ...ice to the load wire s plug socket and plug it into a power source 5 When a power source is applied the voltage electric current and electric power usage will be simultaneously displayed 6 The voltage...

Страница 12: ...ct Power Measurement Warning To prevent an accident during the measurement make sure that the location of HPM 300A is in an appropriate mode before connecting the test lead into the instrument DC load...

Страница 13: ...circuit with measurement load device using crocodile clip Refer to above circuit 5 When a power source is applied the voltage electric current and electric power usage will be simultaneously displayed...

Страница 14: ...measurement range of HPM 300A can be extended from 2 2 kW to 220 kW Warning To prevent an accident during the measurement make sure that the location of HPM 300A is in an appropriate mode before conne...

Страница 15: ...Connect the terminal of an electric current to HPM 300A s Hi Current and the terminal to the 15A terminal 7 When connecting an electric current to a load the electric current value will be displayed 8...

Страница 16: ...16 Clamp Sensor Connections Meathods 1 Single phase two wire electric power measurement 1 2 wire 2 Three phase three wire electric power measurement 3 3 wire www biall com pl...

Страница 17: ...17 3 Three phase four wire electric power measurement 3 4 wire www biall com pl...

Страница 18: ...1 In the Wh mode press ENERGY 2 ENERGY will be displayed on the LCD screen with the electric power usage accumulated will be indicated at the top 3 Press ENERGY to stop the Wh mode 4 To delete the pre...

Страница 19: ...k measurement response or pulse input When the range is set to auto the range may be adjusted frequently if a waveform such as a pulse is input In such a case set the range manually How to set manual...

Страница 20: ...20 DC Voltage Range change LCD View AC Voltage Range change LCD View www biall com pl...

Страница 21: ...w If an input exceeds the maximum permissible RMS limit alarm will sound and OL will be displayed You must increase the manual range or change to auto range in this case AC DC Current manual change LC...

Страница 22: ...ation Method Supported PC program 1 HPM 300A assists RS 232C communication 2 The GUI PC program WattmanViewer will be provided free of charge and an executable file can be downloaded from the homepage...

Страница 23: ...ately 0 5s cycle In a set of data S means the start of the data and E means the end of the data Each item of data to be transmitted is up to six digits The first 5 digits are effective numbers and the...

Страница 24: ...nd Size of byte Function W 1 Command 0 or 1 or 2 1 0 starts the watt hour 1 stops the watt hour 2 resets the accumulated values from the watt hour and the watt hour time into 0 1 Closer Data transmiss...

Страница 25: ...most accurate measurement value can be obtained 3 The value flashed on the screen is the value of the RMS current running through the cable 1A 1mV Caution Measurement cannot be done when more than tw...

Страница 26: ...2 Check the batteries or replace the test lead when necessary 3 Services will be provided free of charge within a year from the purchase date of the equipment 4 When equipment failure occurs due to th...

Отзывы: