32
Using the Software
The Graphic Display
General overview
The graphic display is located in the right main area of the screen.
If an AcQBlate® FORCE ablation catheter is connected, the following
information is displayed graphically depending on the configuration of the
device:
• The angle at which the ablation catheter is applied to the cardiac wall
• The delivery of ablation energy (only if a RF generator is connected to
Qubic Force.)
• Exceedance of the set contact force limit
• A possible foreseeable perforation of the cardiac wall because the
following values are not within the respective tolerance range:
– The contact force is above the set limit.
– And the angle at which the ablation catheter is applied to
the cardiac wall is below the set limit.
The contact force limit (F
max
= 40 g) is preset in the factory settings. To
adjust this value and also set the visual warning limit for the angle at which
the ablation catheter is applied to the cardiac wall, a mouse or keyboard
must be connected and you have to switch to the Settings view (The
Settings View, p. 35).