31
Using the Software
The Numerical Display
General overview
The numerical display is located in the left main area of the screen.
If an AcQBlate® FORCE ablation catheter is connected, the following current
values are shown:
• F: The current contact force of the ablation catheter tip on the cardiac
wall, in grams (g)
• FTI: The current force-time integral in gram seconds (gs)
The force-time integral is calculated from the following formula:
t2
FTI:
F [t] *dt
t1
• t1: Start of radiofrequency ablation
• t2: End or duration since start of radiofrequency ablation
• F: Current contact force
If there is no AcQBlate® FORCE ablation catheter connected, no information
will be displayed in this area.