![Actron E20CP-III Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/actron/e20cp-iii/e20cp-iii_operation-manual_2844044014.webp)
14
Key-in the desired
A
value and depress the [
ENTER
] key to save and move to the
B
variable
The display first shows: <<
CUSTOM – SCALE
>>
QMS = A x (S x S) + B x (S) + C
Followed by the top line of the display showing:
QMS = A x (S x S) + (B x S) + C
alternating with
[
ENTER
]
To CHANGE
alternating with
PRESS
[
FNC
]
FOR NEXT
The bottom line of the display shows:
(B x S) B = .00000
To move to the
C
variable without editing the
B
variable displayed, depress the [
FNC
] key
To edit the
B
variable, depress the [
ENTER
] key
The display will show:
[ENTER] TO SAVE & EXIT
(B x S) B = 00.000
Key-in the desired
B
value and depress the [
ENTER
] key to save and move to the
C
variable
The display first shows: <<
CUSTOM – SCALE
>>
QMS = A x (S x S) + (B x S) + C
Followed by the top line of the display showing:
QMS = A x (S x S) + (B x S) + C
alternating with
[
ENTER
]
To CHANGE
alternating with
PRESS
[
FNC
]
FOR NEXT
The bottom line of the display shows:
(S) +/- C C = +00.00
To move back to the
A
variable without editing the
C
variable displayed, depress the [
FNC
] key
To edit the
C
variable, depress the [
ENTER
] key
Содержание E20CP-III
Страница 25: ...25 XV CE CERTIFICATE DECLARATION of CONFORMITY...