Troubleshooting
Message / Symptom
Possible causes
Solution
“Low Batt”.
Batteries are flat.
Recharge batteries.
“Error 01- 02”
Device is too warm to
Allow device to cool before use.
then shutdown.
operate (above 60C).
“Error 07”
Output shorted (i.e. leads
Check output lead or pads are not
then shutdown.
or electrodes touching)
touching. Turn device off then on.
or safety limit reached.
Otherwise follow instructions for 03-06.
Non-responsive
Internal fault.
Remove and replace batteries.
or locked up display.
Reset device as described on p 31.
Electrodes
Poor adhesion with skin.
Check electrodes are not peeling away.
become hot
during use.
Power set too high.
Reduce power level by 25%.
“Check Leads and Pads”
The electrodes are not
Make sure that the lead wire is properly
on display.
properly stuck to the
connected to both the device and the
skin.
electrodes. Ensure that the electrodes
are firmly stuck to the skin. Replace
Badly connected lead
electrodes if required. If skin is very
wire.
hairy, cut short underneath the
electrodes and try again.
Very hairy skin.
If none of the above works, disable
Pad Detect in Settings (p 32).
Blank screen.
Batteries are flat or not
Recharge batteries and check
correctly inserted.
orientation.
“Error 03-04-05-06”
Internal error.
Turn device off then on, if error persists
then shutdown.
reset device (see instructions on p 31).
“Error 08”
TSE/IC Switch
Check switch is fully over to correct side
then shutdown.
stuck in mid position
Otherwise follow instructions for 03-06.
Display does not
Weak or dead batteries.
Recharge batteries.
come on even
though batteries
Batteries installed the
Check that batteries are inserted the
are inserted.
wrong way around.
right way around.
Beeping noise when
Device not fully pushed
Remove from cradle and reinsert
inserted in cradle
into cradle.
pressing firmly home.