ACS ALCOSIM Скачать руководство пользователя страница 9

INSTRUCTION MANUAL

6

ALCOSIM SETUP AND OPERATION 

STEPS

1. 

With the ALCOSIM breath alcohol simulator power disconnected, turn the 

top housing counter clockwise (left) and detach it from the container. 

2. 

Add the entire contents of the 500 mL bottle of alcohol reference solution 

into the ALCOSIM solution container.

The bottle contains the exact liquid volume required for calibration, which is 

also marked on the ALCOSIM container.

3. 

Return the top housing onto the container, be careful not to over tighten.

4. 

Connect tubing with the round mouthpiece to the ALCOSIM simulator air 

inlet and tubing with the liquid trap mouthpiece to the air outlet.

5. 

Perform a leak check by covering the air outlet tube with your thumb and 

blowing into the air inlet tube.

Air bubbles should not appear in the solution. In case of bubbling, check 

that the ALCOSIM head is tightened properly. Return the ALCOSIM to an 

authorized Service Provider if the problem persists.

6. 

Connect the power cable to the simulator power input and to a wall outlet.

7. 

Switch on the simulator. The following occurs: 

• 

The LCD display illuminates and the simulator beeps

• 

The propeller rotates and the heating element activates

• 

Cold is displayed until solution reaches 32 ºC, wherein the LCD screen 

displays actual solution temperature

• 

The simulator maintains solution at a constant 34 ºC ± 0.02 ºC

Solution heat-up takes about 10 minutes. When 34 ºC ± 0.02 ºC is displayed, 

the simulator is ready to provide a breath sample.

8. 

Use the ALCOSIM simulator to blow air into the breath alcohol tester that 

you are calibrating. There are two ways of doing this:

• 

Manually: take a deep breath and blow for the duration of the breath test 

sample period

• 

Mechanically: use a pump-powered air source (Recommended flow rate: 

14.4 L/min)

Содержание ALCOSIM

Страница 1: ...Instruction manual...

Страница 2: ...tional Inc and are used under license Alcohol Countermeasure Systems is trading style of Alcohol Countermeasure Systems International Inc 2012 2014 Alcohol Countermeasure Systems The information discl...

Страница 3: ...sistance 2 ALCOSIM breath alcohol simulator Overview 2 Safety and precautions 3 ALCOSIM setup usage and disassembly 3 Alcohol reference solution 4 Cleaning the ALCOSIM breath alcohol simulator 4 Calib...

Страница 4: ...r Maintains a precise temperature at 34 C 0 02 C Brushless DC motor operating at 24VDC capable of running 24 7 System for detecting motor failures due to mechanical and electrical defect Audio and vis...

Страница 5: ...te and verify breath alcohol testing devices The simulator is a wet bath calibration verification device FRONT Illuminated ON OFF switch Air outlet tube Baf e plate 500 mL solution fill line Wide view...

Страница 6: ...urface avoid contact with the heating element Place the ALCOSIM simulator on a flat surface free from obstruction Do not expose the simulator to direct sunlight for extended periods of time Do not use...

Страница 7: ...n at room temperature Do not freeze or refrigerate solution Do not ingest solution If solution is ingested do not induce vomiting contact your local poison control centre Keep solution away from eyes...

Страница 8: ...ACS 79 007600 AC power cord part number depends on your region Power supply 24VDC 2 5A ACS 07 000075 50cm of vinyl tubing ACS 70 000002 Liquid Trap Mouthpiece ACS 79 001955 Round mouthpiece ACS 95 00...

Страница 9: ...r in the solution In case of bubbling check that the ALCOSIM head is tightened properly Return the ALCOSIM to an authorized Service Provider if the problem persists 6 Connect the power cable to the si...

Страница 10: ...35 Electric potential meter 2 damage Er 41 Heater malfunction Er 52 The water level in the jar is lower than normal The power supply used is not adequate Fill the jar to the correct liquid volume of...

Страница 11: ...UAL 8 INDEX A Alcohol Reference Solution 4 5 ALCOSIM simulator Parts 2 Product features 1 Setup and operations 6 Assistance 2 C Calibration materials 5 E Error codes 7 S Safety and precautions 3 T Tro...

Страница 12: ...IRM ENG 60 000180 I 2012 2014...

Отзывы: