4

• 

Disconnect the simulator power after use 

• 

CAUTION! After disconnecting the simulator power, allow 10 to 15 minutes for 

the heating element to cool before detaching the top housing 

• 

Empty out the solution container at the end of the work day 

• 

Store the simulator in an environment of 5 to 40 °C and 10 to 85% RH 

ALCOHOL REFERENCE SOLUTION

• 

Use the solution concentration specified in the manual for breath testing 

devices 

• 

Replace solution every 5 days or 20 tests 

• 

Use of solution over time affects alcohol concentration 

• 

Do not use a solution bottle with a broken seal, or an expired bottle 

• 

Never use artificial methods to reheat or cool solution, or the solution 

container 

• 

Keep solution at room temperature 

• 

Do not freeze or refrigerate solution 

• 

Do not ingest solution 

• 

If solution is ingested, do not induce vomiting; contact your local poison 

control centre 

• 

Keep solution away from eyes 

• 

If solution comes in contact with eyes, flush eyes with water; if irritation 

continues, contact your local poison control centre 

• 

Refer to your local environmental regulations for more information about safe 

solution-disposal amounts 

• 

Dispose solution down a drain

CLEANING THE ALCOSIM BREATH ALCOHOL 

SIMULATOR

• 

Disconnect the ALCOSIM breath alcohol simulator power before cleaning 

• 

CAUTION! After disconnecting the ALCOSIM simulator power, allow 10 to 15 

minutes for the heating element to cool off before detaching the top housing 

• 

Do not submerge the top housing in water 

• 

Clean the top housing by wiping with a water dampened cloth 

• 

Wash the solution container with plain water and dry with paper towel

Содержание ALCOSIM

Страница 1: ...Instruction manual...

Страница 2: ...tional Inc and are used under license Alcohol Countermeasure Systems is trading style of Alcohol Countermeasure Systems International Inc 2012 2014 Alcohol Countermeasure Systems The information discl...

Страница 3: ...sistance 2 ALCOSIM breath alcohol simulator Overview 2 Safety and precautions 3 ALCOSIM setup usage and disassembly 3 Alcohol reference solution 4 Cleaning the ALCOSIM breath alcohol simulator 4 Calib...

Страница 4: ...r Maintains a precise temperature at 34 C 0 02 C Brushless DC motor operating at 24VDC capable of running 24 7 System for detecting motor failures due to mechanical and electrical defect Audio and vis...

Страница 5: ...te and verify breath alcohol testing devices The simulator is a wet bath calibration verification device FRONT Illuminated ON OFF switch Air outlet tube Baf e plate 500 mL solution fill line Wide view...

Страница 6: ...urface avoid contact with the heating element Place the ALCOSIM simulator on a flat surface free from obstruction Do not expose the simulator to direct sunlight for extended periods of time Do not use...

Страница 7: ...n at room temperature Do not freeze or refrigerate solution Do not ingest solution If solution is ingested do not induce vomiting contact your local poison control centre Keep solution away from eyes...

Страница 8: ...ACS 79 007600 AC power cord part number depends on your region Power supply 24VDC 2 5A ACS 07 000075 50cm of vinyl tubing ACS 70 000002 Liquid Trap Mouthpiece ACS 79 001955 Round mouthpiece ACS 95 00...

Страница 9: ...r in the solution In case of bubbling check that the ALCOSIM head is tightened properly Return the ALCOSIM to an authorized Service Provider if the problem persists 6 Connect the power cable to the si...

Страница 10: ...35 Electric potential meter 2 damage Er 41 Heater malfunction Er 52 The water level in the jar is lower than normal The power supply used is not adequate Fill the jar to the correct liquid volume of...

Страница 11: ...UAL 8 INDEX A Alcohol Reference Solution 4 5 ALCOSIM simulator Parts 2 Product features 1 Setup and operations 6 Assistance 2 C Calibration materials 5 E Error codes 7 S Safety and precautions 3 T Tro...

Страница 12: ...IRM ENG 60 000180 I 2012 2014...

Отзывы: