22. Connect the back of the HS to the right-side HS fastening pad, with the HS sampling port facing forward.
23. Switch the CS to
“Pump Enabled”
(front)
.
24. Press the ECU button to
Zero
tests.
NOTE:
The test counter on the ECU counts down from 25 tests/5 days
Change the solution after 25 tests or 5 days (whichever comes first)
Use of the solution beyond 25 tests will lead to
out of range
/
inconsistent calibration
action by the CS, and cause
return of the HS
8.2
Functional test
When an operational HS is connected, the ECU initially displays the old calibration factor, followed by
WAIT
. It then begins a
warming up and functional test process which checks the HS for the following:
Blow tube temperature
Pump temperature
Fuel cell temperature
Pump operation
ATTENTION!
Failure of any of these tests will cause the HS to be disabled, and no further testing can be undertaken
until the unit has been returned to your distributor. If a defective HS is connected, the ECU will display the error
message and make a rejection sound, and prevent testing.
On successful completion of the functional testing, the ECU will display
PRESS BUTTON TO START
to subject the HS to
verification and calibration processes.
ATTENTION!
A calibration cannot commence unless the ALCOSIM communication cable is connected and the alcohol
solution is at the correct operating temperature.