Ultrasonic velocity tester A1410 Pulsar
Operation Manual
28
ACOUSTIC
CONTROL
SYSTEMS
2
2.3.2.2 Saving the results
The measurement results are saved in the instrument memory in cells. Cells form groups. Groups and cells in the groups are
identified by index numbers. Groups and cells in the groups are numbered beginning with one.
The maximum number of cells in a group is 500.
The maximum number of groups is 100.
When the maximum number of cells is reached in a group, the appropriate information is indicated on the screen of the
instrument.
According to practice it is more com-
fortable to record the results in small groups
consisting of several dozens of measure-
ments. If necessary, you can return to any
existing group and continue recording the
results in that group.
To save a measurement please select
any existing group or create a new one be-
fore you start the measurements. After the
measurement process is finished please
press ENTER. The measurement results will
be saved in the first empty cell of the current
(selected) group.
Note:
when the measurement result
does not conform with the permitted range,
the measurement is regarded as incorrect.
When you try to save the following window
will be indicated on the screen (Figure 29).
Figure 29
Содержание A1410 PULSAR
Страница 2: ...Ultrasonic velocity tester A1410 Pulsar Operation Manual 2 ACOUSTIC CONTROL SYSTEMS ...
Страница 37: ...Ultrasonic velocity tester A1410 Pulsar Operation Manual 37 ACOUSTIC CONTROL SYSTEMS NOTES ...
Страница 38: ...Ultrasonic velocity tester A1410 Pulsar Operation Manual 38 ACOUSTIC CONTROL SYSTEMS NOTES ...
Страница 39: ...Ultrasonic velocity tester A1410 Pulsar Operation Manual 39 ACOUSTIC CONTROL SYSTEMS ...