Ultrasonic flaw detector А1214 EXPERT
Operation Manual
49
ACOUSTIC
CONTROL
SYSTEMS
Table 21
Key
Function
Move the strobe in the appropriate direction.
When moving the multilevel strobe in vertical direction, the acceptance level will be moved; at that the
reference and search levels will be moved relative to the acceptance level according to the settings
Change the length of the strobe relative to its left boundary
Switch Off the strobe and exit the settings.
To switch Off the strobe, enter the settings of the corresponding strobe by means of the F1 or F2
keys correspondingly
F1 or F2 exit the editing mode
F2 or F1. F3 – F5 exit the editing mode, functions of the enabled key will be performed
Enter the SETTING mode
the maximum), corresponding parameters of the cursor in this point will be indicated on the screen; sound and light indication
is produced (AFAS actuates).
If the strobe is beyond the range displayed on the screen, then a pointer will appear at the right boundary of the range. The
pointer allows determination of the strobe threshold level (a single red arrow for the first strobe and a double blue arrow for
the second strobe).
ATTENTION: IF THE SIGNAL AMPLITUDE EXCEEDS THE LEVEL OF THE STROBE RANGE DISPLAYED ON THE SCREEN, THEN
AFAS WILL BE ACTUATED BUT THE CURSOR AND THE MARKER WILL NOT BE DISPLAYED!
Each strobe has two modes: a single-level and a multilevel strobe. The multilevel strobe allows simultaneous setting of
three sensitivity levels: acceptance, reference and search.
ATTENTION: IF THE REFERENCE LEVEL IS ON, THEN THE STROBE CANNOT BE MOVED IN VERTICAL DIRECTION!
The functions of the keys applicable for active icon
or
for the first strobe, and
or
for the second strobe, are presented in the Table 21.
2
Содержание A1214 Expert
Страница 2: ...Ultrasonic flaw detector А1214 EXPERT Operation Manual 2 ACOUSTIC CONTROL SYSTEMS ...
Страница 61: ...Ultrasonic flaw detector А1214 EXPERT Operation Manual 61 ACOUSTIC CONTROL SYSTEMS NOTES ...
Страница 62: ...Ultrasonic flaw detector А1214 EXPERT Operation Manual 62 ACOUSTIC CONTROL SYSTEMS NOTES ...
Страница 63: ...Ultrasonic flaw detector А1214 EXPERT Operation Manual 63 ACOUSTIC CONTROL SYSTEMS ...