
Ultrasonic flaw detector А1214 EXPERT
Operation Manual
33
ACOUSTIC
CONTROL
SYSTEMS
2
Figure 27
Figure 28
- find the maximal signal from the second reference reflector and adjust it in the same manner, at that the first anchor point
shall be out of the strobe field (Figure 27);
- press the
key, at that the second anchor point will appear on the peak value of the signal;To delete the anchor point,
press the F6
key.
Note:
several anchor points will be deleted according to their creation sequence.
- create the anchor points for all reference reflectors;
- press the
key;
- confirm or cancel the DAC curve (Figure 28).
Press the
key to exit the configuration window not saving the settings.
Содержание A1214 Expert
Страница 2: ...Ultrasonic flaw detector А1214 EXPERT Operation Manual 2 ACOUSTIC CONTROL SYSTEMS ...
Страница 61: ...Ultrasonic flaw detector А1214 EXPERT Operation Manual 61 ACOUSTIC CONTROL SYSTEMS NOTES ...
Страница 62: ...Ultrasonic flaw detector А1214 EXPERT Operation Manual 62 ACOUSTIC CONTROL SYSTEMS NOTES ...
Страница 63: ...Ultrasonic flaw detector А1214 EXPERT Operation Manual 63 ACOUSTIC CONTROL SYSTEMS ...